File name: MAN.MAS.EN‐‐02X‐‐Sherlog‐Master‐user‐manual‐Eng.docx
31/60
11.3
SDT 8 MS transmitter description
The on‐off switch
Located at the base, this switch (3) enables the equipment to be switched on (ON position) or off (OFF position).
The 6 position multi‐setting potentiometer
It regulates the total output power of the ultrasonic signal; level 1 provides the minimum level and 6 the maximum.
The charge connector
Marked LOAD, the charge connector enables to re‐load the internal battery
without removing it. Overcharging is not possible when used with the SDT
charger provided. Batteries can be charged during use. (Batteries will charge
slower when the equipment is in use)
The charge connector is located
on the side of the multi‐transmitter.
The battery
The sealed lead‐acid gel type battery is accessible after
having removed the bottom protection plate. The rated
voltage is 12V and its capacity is 1.2Ah at 20 °C; the
autonomy is 2.5 to 3.5 hour for a full charge. The position of
the battery in its compartment is of no importance from a
polarity point of view.
The multi‐transmitter serial number
It is located at the bottom, on the removable plate.
The battery is accessible from the bottom
Localization of the serial number
of the multi‐transmitter.
Summary of Contents for Sherlog Kit Master SDT270
Page 6: ...File name MAN MAS EN 02X Sherlog Master user manual Eng docx 6 60 First and foremost ...
Page 10: ...File name MAN MAS EN 02X Sherlog Master user manual Eng docx 10 60 The SDT270 receiver ...
Page 29: ...File name MAN MAS EN 02X Sherlog Master user manual Eng docx 29 60 The SDT8 MS transmitter ...
Page 50: ...File name MAN MAS EN 02X Sherlog Master user manual Eng docx 50 60 Technical specifications ...
Page 54: ...File name MAN MAS EN 02X Sherlog Master user manual Eng docx 54 60 Appendixes ...