SECTION 6 MAINTENANCE
PDS-2010
dPOS - Decimal Point Position
This parameter, applicable only if a linear input is fitted, defines the position of the decimal point in
values of the process variable, set point, alarm levels, and recorder output as follows:
VALUE
DECIMAL POINT POSITION
0
XXXX
1
XXX.X
2
XX.XX
3
X.XXX
Filt - Input Filter Time Constant
The controller input is equipped with a digital filter that is used to filter out any extraneous impulses on
the process variable. This filtered process variable is used for all process variable dependent functions.
The time constant for this filter may be adjusted in the range 0.0 seconds to 100.0 seconds in .5 second
increments.
NOTICE: If this parameter is set to an excessively high value, the control quality may be
significantly impaired. Choose a value that is sufficiently large to attenuate stray noise on the
process variable signal, but no larger.
Inhi - Alarm Inhibit
This inhibits an alarm at power-up, or if a controller set point is switched, until that alarm goes inactive.
The alarm operates normally from that point onwards.
InPt - Input Range
This is the overall input range and type as selected by the Inpt parameter in Configuration Mode.
LAEn - Loop Alarm
This is a controller parameter; it is the means by which the user can enable or disable the loop alarm. The
loop alarm is a special alarm that detects faults in the control feedback loop by continuously monitoring
process variable response to the control output(s). The loop alarm can be tied to any suitable output.
When enabled, the loop alarm repeatedly checks if the control output(s) are at the maximum or minimum
limit. If an output is at the limit, an internal timer is started. Thereafter, if the high output has not caused
the process variable to be corrected by a predetermined amount 'V' after time 'T' has elapsed, the loop
alarm becomes active. Subsequently, the loop alarm mode repeatedly checks the process variable and the
control output(s). When the process variable starts to change value in the correct sense or when the output
is no longer at the limit, the loop alarm is deactivated.
OFFS - Process Variable Offset
This parameter is used to modify the actual process variable value (measured at the Controller’s input
terminals) in the following manner:
Offset Process Variable Value = Actual Process Variable Value + Process Variable Offset Value.
OPuL - Output 1 Power Limit
This parameter is used to limit the power level of Output 1 and may be used to protect the process being
controlled. If no protection is required, this parameter may be set to 100%. The parameter may be
adjusted between 0% and 100%.
Summary of Contents for LABCOTER 2
Page 3: ...OM 610 1002 1 Operator s Manual Rev 37...
Page 4: ...Rev 37 Operator s Manual OM 610 1002 1...
Page 14: ...Rev 37 Operator s Manual OM 610 1002 1 page 6 SECTION 2 OVERVIEW PDS 2010...
Page 44: ......
Page 66: ...Rev 37 Operator s Manual OM 610 1002 1 page 58 SECTION 6 MAINTENANCE PDS 2010...
Page 68: ...Rev 37 Operator s Manual OM 610 1002 1 page 60 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 70: ...Rev 37 Operator s Manual OM 610 1002 1 page 62 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 72: ...Rev 37 Operator s Manual OM 610 1002 1 page 64 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 74: ...Rev 37 Operator s Manual OM 610 1002 1 page 66 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 76: ...Rev 37 Operator s Manual OM 610 1002 1 page 68 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 78: ...Rev 37 Operator s Manual OM 610 1002 1 page 70 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 80: ...Rev 37 Operator s Manual OM 610 1002 1 page 72 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 82: ...Rev 37 Operator s Manual OM 610 1002 1 page 74 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 84: ...Rev 37 Operator s Manual OM 610 1002 1 page 76 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 86: ...Rev 37 Operator s Manual OM 610 1002 1 page 78 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 88: ...Rev 37 Operator s Manual OM 610 1002 1 page 80 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 90: ...Rev 37 Operator s Manual OM 610 1002 1 page 82 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 92: ...Rev 37 Operator s Manual OM 610 1002 1 page 84 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 94: ...Rev 37 Operator s Manual OM 610 1002 1 page 86 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 96: ...Rev 37 Operator s Manual OM 610 1002 1 page 88 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 98: ...Rev 37 Operator s Manual OM 610 1002 1 page 90 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 100: ...Rev 37 Operator s Manual OM 610 1002 1 page 92 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 102: ...Rev 37 Operator s Manual OM 610 1002 1 page 94 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 104: ...Rev 37 Operator s Manual OM 610 1002 1 page 96 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 106: ...Rev 37 Operator s Manual OM 610 1002 1 page 98 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 108: ...Rev 37 Operator s Manual OM 610 1002 1 page 100 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 110: ...Rev 37 Operator s Manual OM 610 1002 1 page 102 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 112: ...Rev 37 Operator s Manual OM 610 1002 1 page 104 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 114: ...Rev 37 Operator s Manual OM 610 1002 1 page 106 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 116: ...Rev 37 Operator s Manual OM 610 1002 1 page 108 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 128: ...Rev 37 Operator s Manual OM 610 1002 1 page 120 SECTION 10 WARRANTY PDS 2010...
Page 130: ...Rev 37 Operator s Manual OM 610 1002 1 page 122 SECTION 10 WARRANTY PDS 2010...
Page 131: ...PDS 2010 SCHEMATICS page 123 ES 610 1001 2 CE ELECTRICAL SCHEMATIC 1 OF 2...
Page 132: ...page 124 SCHEMATICS PDS 2010...
Page 133: ...PDS 2010 SCHEMATICS page 125 ES 610 1001 2 CE ELECTRICAL SCHEMATIC 2 OF 2...
Page 134: ...page 126 SCHEMATICS PDS 2010...
Page 135: ...PDS 2010 SCHEMATICS page 127 ES 610 1068 1 DOMESTIC ELECTRICAL SCHEMATIC 1 OF 3...
Page 136: ...page 128 SCHEMATICS PDS 2010...
Page 137: ...PDS 2010 SCHEMATICS page 129 ES 610 1068 1 DOMESTIC ELECTRICAL SCHEMATIC 2 OF 3...
Page 138: ...page 130 SCHEMATICS PDS 2010...
Page 139: ...PDS 2010 SCHEMATICS page 131 ES 610 1068 1 DOMESTIC ELECTRICAL SCHEMATIC 3 OF 3...
Page 140: ...page 132 SCHEMATICS PDS 2010...
Page 141: ...PDS 2010 SCHEMATICS page 133 ES 610 1001 5 220V ELECTRICAL SCHEMATIC 1 OF 3...
Page 142: ...page 134 SCHEMATICS PDS 2010...
Page 143: ...PDS 2010 SCHEMATICS page 135 ES 610 1001 5 220V ELECTRICAL SCHEMATIC 2 OF 3...
Page 144: ...page 136 SCHEMATICS PDS 2010...
Page 145: ...PDS 2010 SCHEMATICS page 137 ES 610 1001 5 220V ELECTRICAL SCHEMATIC 3 OF 3...
Page 146: ...page 138 SCHEMATICS PDS 2010...
Page 147: ...PDS 2010 SCHEMATICS page 139 ES 610 1065 1 200V ELECTRICAL SCHEMATIC 1 OF 3...
Page 148: ...page 140 SCHEMATICS PDS 2010...
Page 149: ...PDS 2010 SCHEMATICS page 141 ES 610 1065 1 200V ELECTRICAL SCHEMATIC 2 OF 3...
Page 150: ...page 142 SCHEMATICS PDS 2010...
Page 151: ...PDS 2010 SCHEMATICS page 143 ES 610 1065 1 200V ELECTRICAL SCHEMATIC 3 OF 3...
Page 152: ...page 144 SCHEMATICS PDS 2010...