Rev 37
Operator’s Manual
OM-610-1002-1
page 14
SECTION 5 OPERATION
PDS-2010
TABLE 5-1
Equivalence Table for Pressure/Vacuum Measurement
Vacuum Torr milliTorr PSIA
Inches
Hg
Atmosphere
Atmosphere 760
760
000 14.7
29.92
1.0
Low 100
100
000
1.93 3.94
0.132
Low 1 1000
0.019
0.0394
0.00132
Med. 0.100 100
1.93x10-3 3.94x10-3 1.32x10-4
Med. 0.010 10 1.93x10-4 3.94x10-4 1.32x10-5
High
0.001 = (10
-3
) 1 1.93x10-5 3.94x10-5 1.32x10-6
Very High
(10
-6
) 0.001
(10
-3
)
1.93x10-8 3.94x10-8 1.32x10-9
Ultra High
(10
-9
) (10
-6
)
1.93x10-11 3.94x10-11 1.32x10-12
5.2.3 Vacuum Levels
Vacuum is usually divided into five levels, qualitatively labeled: Low, Medium, High, Very High, and Ultra
High. A Low vacuum related to having a relatively high pressure and a Very High vacuum having a relatively
low pressure.
Low Vacuum
represents pressure above 1 Torr absolute. Flow in this range is viscous, as represented by most
common fluids. Mechanical vacuum pumps are used for low vacuum, and represent the large majority of
pumps in industrial practice.
Medium Vacuum
represents pressures between 1 and 1x10
-3
Torr absolute. This is a transition range between
viscous and molecular flow. Most pumps serving this range are also mechanical. The 2010 performs the
Parylene deposition process in the medium vacuum range.
High Vacuum
represents pressures between 1x10
-3
and 1x10
-6
-Torr absolute. Flow in this region is molecular,
with very little interaction between individual molecules. Many specialized applications for the semiconductor
manufacturing industry fall in this range. Diffusion pumps, turbomolecular, and cryopumps are typically used
in this range.
Very High Vacuum
represents absolute pressures between 1x10
-6
and 1x10
-9
Torr.
Ultra High Vacuum
represents absolute pressure below 1x10
-9
Torr.
5.2.4 Vacuum Pumps
A vacuum pump is a device that is used to reduce and maintain the gas pressure in a vessel below atmospheric
pressure. The 2010 Parylene Deposition System uses an EDWARDS RV 8 (60 HZ) or EDWARDS RV12 (50
HZ) vacuum pump.
5.2.5 Gas Flow
When a closed container at atmospheric pressure is evacuated by virtue of a vacuum pumping system, the
flow of gas through the container will pass through one or more flow regimes. The flow regime in a container
Summary of Contents for LABCOTER 2
Page 3: ...OM 610 1002 1 Operator s Manual Rev 37...
Page 4: ...Rev 37 Operator s Manual OM 610 1002 1...
Page 14: ...Rev 37 Operator s Manual OM 610 1002 1 page 6 SECTION 2 OVERVIEW PDS 2010...
Page 44: ......
Page 66: ...Rev 37 Operator s Manual OM 610 1002 1 page 58 SECTION 6 MAINTENANCE PDS 2010...
Page 68: ...Rev 37 Operator s Manual OM 610 1002 1 page 60 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 70: ...Rev 37 Operator s Manual OM 610 1002 1 page 62 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 72: ...Rev 37 Operator s Manual OM 610 1002 1 page 64 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 74: ...Rev 37 Operator s Manual OM 610 1002 1 page 66 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 76: ...Rev 37 Operator s Manual OM 610 1002 1 page 68 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 78: ...Rev 37 Operator s Manual OM 610 1002 1 page 70 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 80: ...Rev 37 Operator s Manual OM 610 1002 1 page 72 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 82: ...Rev 37 Operator s Manual OM 610 1002 1 page 74 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 84: ...Rev 37 Operator s Manual OM 610 1002 1 page 76 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 86: ...Rev 37 Operator s Manual OM 610 1002 1 page 78 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 88: ...Rev 37 Operator s Manual OM 610 1002 1 page 80 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 90: ...Rev 37 Operator s Manual OM 610 1002 1 page 82 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 92: ...Rev 37 Operator s Manual OM 610 1002 1 page 84 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 94: ...Rev 37 Operator s Manual OM 610 1002 1 page 86 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 96: ...Rev 37 Operator s Manual OM 610 1002 1 page 88 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 98: ...Rev 37 Operator s Manual OM 610 1002 1 page 90 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 100: ...Rev 37 Operator s Manual OM 610 1002 1 page 92 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 102: ...Rev 37 Operator s Manual OM 610 1002 1 page 94 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 104: ...Rev 37 Operator s Manual OM 610 1002 1 page 96 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 106: ...Rev 37 Operator s Manual OM 610 1002 1 page 98 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 108: ...Rev 37 Operator s Manual OM 610 1002 1 page 100 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 110: ...Rev 37 Operator s Manual OM 610 1002 1 page 102 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 112: ...Rev 37 Operator s Manual OM 610 1002 1 page 104 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 114: ...Rev 37 Operator s Manual OM 610 1002 1 page 106 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 116: ...Rev 37 Operator s Manual OM 610 1002 1 page 108 SECTION 7 COMPONENT LOCATION PDS 2010...
Page 128: ...Rev 37 Operator s Manual OM 610 1002 1 page 120 SECTION 10 WARRANTY PDS 2010...
Page 130: ...Rev 37 Operator s Manual OM 610 1002 1 page 122 SECTION 10 WARRANTY PDS 2010...
Page 131: ...PDS 2010 SCHEMATICS page 123 ES 610 1001 2 CE ELECTRICAL SCHEMATIC 1 OF 2...
Page 132: ...page 124 SCHEMATICS PDS 2010...
Page 133: ...PDS 2010 SCHEMATICS page 125 ES 610 1001 2 CE ELECTRICAL SCHEMATIC 2 OF 2...
Page 134: ...page 126 SCHEMATICS PDS 2010...
Page 135: ...PDS 2010 SCHEMATICS page 127 ES 610 1068 1 DOMESTIC ELECTRICAL SCHEMATIC 1 OF 3...
Page 136: ...page 128 SCHEMATICS PDS 2010...
Page 137: ...PDS 2010 SCHEMATICS page 129 ES 610 1068 1 DOMESTIC ELECTRICAL SCHEMATIC 2 OF 3...
Page 138: ...page 130 SCHEMATICS PDS 2010...
Page 139: ...PDS 2010 SCHEMATICS page 131 ES 610 1068 1 DOMESTIC ELECTRICAL SCHEMATIC 3 OF 3...
Page 140: ...page 132 SCHEMATICS PDS 2010...
Page 141: ...PDS 2010 SCHEMATICS page 133 ES 610 1001 5 220V ELECTRICAL SCHEMATIC 1 OF 3...
Page 142: ...page 134 SCHEMATICS PDS 2010...
Page 143: ...PDS 2010 SCHEMATICS page 135 ES 610 1001 5 220V ELECTRICAL SCHEMATIC 2 OF 3...
Page 144: ...page 136 SCHEMATICS PDS 2010...
Page 145: ...PDS 2010 SCHEMATICS page 137 ES 610 1001 5 220V ELECTRICAL SCHEMATIC 3 OF 3...
Page 146: ...page 138 SCHEMATICS PDS 2010...
Page 147: ...PDS 2010 SCHEMATICS page 139 ES 610 1065 1 200V ELECTRICAL SCHEMATIC 1 OF 3...
Page 148: ...page 140 SCHEMATICS PDS 2010...
Page 149: ...PDS 2010 SCHEMATICS page 141 ES 610 1065 1 200V ELECTRICAL SCHEMATIC 2 OF 3...
Page 150: ...page 142 SCHEMATICS PDS 2010...
Page 151: ...PDS 2010 SCHEMATICS page 143 ES 610 1065 1 200V ELECTRICAL SCHEMATIC 3 OF 3...
Page 152: ...page 144 SCHEMATICS PDS 2010...