48
CHAPTER 5: OPERATION
087-0048 REV B
SCOTT SAFETY
Powering Up the Test
Station
This section describes the power up sequence.
details the sequence.
The power supply adaptor functions as the main power disconnect.
Hibernating the
Monitor Using the Test
Station
This section describes hibernating the multiple monitors using the Test Station. This
only applies to specific monitors designed to hibernate.
1
Ensure the Test Station is powered and a USB Memory is installed.
2
Place the desired number of Monitors into the Test Station. The Test Station may
hibernate up to four (4) Monitors at once.
3
Press and Hold both the BUMP TEST and CALIBRATION buttons for about two
(2) seconds.
Allow Hibernate must be checked in the Test Station Software and a WRITE command
performed prior to pushing the BUMP and CALIBRATION buttons for this method to
be used.
4
A successful hibernation results in a GREEN light corresponding to the Monitor’s
Unit Number.
The Event Log is automatically stored onto the Test Station’s USB Memory.
5
Confirm the Monitor’s LCD is blank.
6
If you encounter any problems, please contact us.
Table 19
Test Station Power-Up Sequence
ACTION
RESULTS
Press the
Bump Test
button.
The Test Station starts and runs through a self-test.
During the self-test ensure the following:
•
The Power LED & the four (4) Unit LEDs lite Orange
•
The Power LED lites Green
•
The four (4) Unit LEDs lite Red
•
The four (4) Unit LEDs turn off & the Power LED remains Green
Note: The Test Station automatically powers off after 120 seconds, and the Power
LED turns off.
Summary of Contents for Protege ZM
Page 1: ...Protege ZM System Guide PROTEGE ZM PORTABLE GAS SYSTEM 087 0048 Rev B ...
Page 2: ......
Page 3: ...Protege ZM System Guide PROTEGE ZM PORTABLE GAS SYSTEM 087 0048 Rev B ...
Page 10: ...X LIST OF TABLES 087 0048 REV B SCOTT SAFETY ...
Page 20: ...XX ABOUT THIS GUIDE 087 0048 REV B SCOTT SAFETY ...
Page 24: ...4 CHAPTER 1 QUICK REFERENCE 087 0048 REV B SCOTT SAFETY ...
Page 30: ...10 CHAPTER 2 INTRODUCTION 087 0048 REV B SCOTT SAFETY ...
Page 44: ...24 CHAPTER 3 INSTALLATION 087 0048 REV B SCOTT SAFETY ...
Page 62: ...42 CHAPTER 4 CONFIGURATION AND SETUP 087 0048 REV B SCOTT SAFETY ...
Page 98: ...78 CHAPTER 6 MAINTENANCE 087 0048 REV B SCOTT SAFETY ...
Page 114: ...94 INDEX 087 0048 REV B SCOTT SAFETY ...
Page 115: ......