15.
C
OMMUNICATION INTERFACES
USB INTERFACE
The USB interface is
the preferred interface for the first configuration of the device. It is the only interface which can never be disabled. To access
it,
NXperience
softw
are for Windows needs to be installed. The USB driver will thus be installed (see the
Configuration Software
ch
apter). A
standard micro-USB cable (not supplied) must be used for desktop or notebook connection.
When connecting the USB cable, the respective icon should light up on the device’s display, indicating that its interface is ready for use. For the first
use, the user will need to wait for Windows to automatically install the driver that comes pre-installed by
NXperience
.
NXperience
will c
onfigure,
monitor and collect tags from the device via USB interface.
The USB interface IS NOT ISOLATED.
It is intended for te
mporary use during CONFIGURATION, MONITORING and
LOG DOWNLOAD. For the safety of people and devices, it should only be
used when the device is disconnected from the external power supply.
The USB interface
can be used in any other connection condition, although
the decision requires careful analysis by the person in charge of
installation.
Fig. 44
– USB Interface
CELLULAR INTERFACE
Fig. 45
– Cellular and GPS Interface
55/66