Hydrim L110w Service Guide 95-109977-1
…
7
Electrical Connection
This appliance must be correctly grounded! The manufacture cannot be held responsible for damage or
injury caused by incorrect or missing grounding. Before making any connections check that the voltage
shown on the serial number label corresponds to your power supply. The machine is supplied as standard
for connection to 208-240V 60 Hz single-phase power supply and is fitted with a power supply cord
1.8m/6ft. long, with a 6-15 NEMA plug. It should be connected to the main power supply according to the
information below.
Voltage:
208-240V
Frequency:
60 Hz
Rated load:
2.5 kW
Circuit breaker:
15A
If you need to extend the mains connecting lead, do not cut off the plug. Replace the whole cable, which
can be disconnected inside the unit.
Connection to the Water Intake
The unit must be connected to the water supply in accordance with all local and national plumbing codes.
SciCan recommends a hard plumbing installation within 1.5m/5ft. of the unit. If additional distance is
necessary, commercial grade plumbing hose must be used to minimize leaks.
Connect inlet hoses to hot and cold water taps using the hoses connected to the unit and in accordance with
the installation instructions.
Water Connection
Water Pressure:
1-10 bar/14.5-145 psi
Water Temperature:
Cold water less than 30ºC/86ºF
Hot water up to 60ºC/140ºF
Drainage
The unit is supplied with a 1.5m/5ft. flexible drain hose with a 2cm/3/4” barb inlet. The hose should not be
shortened or attached to any fittings that would cause a reduction in water flow. The drain system is
equipped with a non-return valve that prevents dirty water from flowing back into the unit.
The drain hose should not be further than 1.5m/5ft. from a hard plumbing drain. If this is not possible, then
commercial grade plumbing hose must be used to minimize leaks.
The hose can be attached to an existing drain through the use of a 3.5cm/11/2” or larger stand pipe/ P-trap
combination. Alternatively, the hose can be connected directly to the existing drain lines, provided the
fittings or adapters used do not reduce the water flow. The drain hose should not exceed 3.3m/13ft. in
length, or be attached to the main drain at a point higher the 35cm/14” above the floor.