Hydrim L110w Service Guide 95-109977-1
…
24
Press the back arrow to return to the previous menu.
Shows the status of each component in
the unit:
Chamber Full SW – ON/OFF
Salt (Regeneration) SW – OK
Rinse Aid SW – ON/OFF
Chemical Sensor - ON/OFF
Chamber Overflow SW – ON/OFF
Chamber Pressure SW – ON/OFF
Door SW – ON/OFF
Chamber Temperature - XXºC
CTS (clear to send) – ON/OFF
(To Printer or Datalogger)
Press the back arrow to return to the
previous menu.
__________________________________________________________________________________
Allows individual testing of the following
components:
Circ. Pump M1 - ON/OFF
All Devices – ON/OFF
Latch L1 – ON/OFF
Salt Valve Y2 – ON/OFF
Dosing Pump M4 – ON/OFF
Dryer Motor+Heater – ON/OFF
Hot W Valve Y7 – ON/OFF
Cold W Valve Y1 – ON/OFF
Rinse Aid Valve – ON/OFF
Waste Pump M2 – ON/OFF
Up arrow pressed, next device.
Down arrow pressed, previous device.
On button pressed, set device On.
Off button pressed, set device Off.
Press the back arrow to return to the previous menu.
__________________________________________________________________________________
Up arrow button pressed, increase
value.
Down arrow button pressed, decrease
value.
Default button pressed, set to default
value.
Press the back arrow to return to the
previous menu.
Diagnostic Tools
Diagnostic Tools
Component Test
Error History
Test Cycle
Component Test
Diagnostic Tools
Set Debug Screen
Circ. Pump M1 OFF
Off
On
View IO Status
Chamber Full SW OFF
Salt SW OK
Rinse aid OFF
Chemical Sensor OFF
Chamber Overflow OFF
Chamber Pressure OFF
Door SW ON
Chamber T 27ºC
CTS ON
Error History
View IO Status
Component Test
Set debug screen
Off
Default: Off