SciCan HYDRIM L110w Service Manual Download Page 36

 

Hydrim L110w Service Guide 95-109977-1

    

36 

 
Safety Switch & Probe 

 

Remove filter screen from inside 
chamber and unclip Safety 
Switch Probe from Heating 
Element. 
Behind the cover below the front 
door remove the nut that holds 
the probe to the chamber. Safety 
Switch is located behind plate. 

Safety Switch Part # 22815.03

 

 

 
Chamber Temperature Sensor 
 

Remove mounting nut from 
sensor located underneath the 
filter inside the chamber. Behind 
the cover below the front door 
remove the sensor from the 
chamber and unplug the red 
wiring. 

Chamber Temp. 

Sensor Part # 22816.12

  

 
Door Latch 
 

Remove top cover. Disconnect Door Latch in front of unit and 
replace. 

Door Latch Part # 23066.00

  

 
 
 
 
 
 
 
 

Touch Screen 

 

Remove the screw above the touch screen and open door. Disconnect I/O Board 
and Logic board from Touchscreen, replace the touchscreen and re-connect the 
boards in reverse order. Close the door and re-insert the screw above the 
touchscreen. 

Touch Screen Part # 01-109785S  
 

 
 
 
 
 

Probe mounting nut 

Safety Switch location  

Safety Switch 

Chamber Temperature Sensor 

 

Chamber Temperature Sensor 

 

Summary of Contents for HYDRIM L110w

Page 1: ...Service Manual HYDRIM L110w Hydrim L110w Service Guide 95 109977 Rev 1 0 Copyright 2007 SciCan Ltd All rights reserved...

Page 2: ...iew of Menus 13 User Menu 14 15 Setup Menu 16 18 Technical Service Menu Overview 19 21 Technical Service Menu 22 29 Troubleshooting Using Technical Service Menu 30 Cycle Faults 31 33 Opening Door in C...

Page 3: ...ng a rinse and hold cycle Cycles times depend on the temperature of incoming water Water consumption depends on the regeneration Level Specifications Height freestanding 850 mm 33 5 Height built in 83...

Page 4: ...actor as it is an Installation Category 2 device SciCan shall not be liable for incidental special or consequential damages caused by any maintenance or service performed on the Hydrim L110w by a thir...

Page 5: ...CRIPTION DESCRIPTION Nose pliers Allen key 3 0 mm Screwdriver PH 1 Channel Locks Screwdriver PH 2 Screwdriver Slot Wire cutters Small Slot screwdriver Nut driver 8 mm Shipping Instructions The unit sh...

Page 6: ...drain hoses ensure that you use commercial grade plumbing hose The maximum length of the drain hose is 3 3 m 13 Installation Instructions The Hydrim L110w should only be installed and service by a qua...

Page 7: ...ance is necessary commercial grade plumbing hose must be used to minimize leaks Connect inlet hoses to hot and cold water taps using the hoses connected to the unit and in accordance with the installa...

Page 8: ...idual strip packages Take the water sample from the location where the machine will be installed Once you have determined the water hardness level in ppm parts per million find the recommended setting...

Page 9: ...il it is full with water It is not necessary to add water during subsequent refills of the salt container Replace the salt container lid After salt has been added to the unit the softening salt indica...

Page 10: ...ockwise lower is clockwise and remove arm Inspect nozzles for debris rinse under tap and reassemble Yearly Water Inlet Hoses Disconnect the hot and cold water inlet hoses Make sure the hoses are clean...

Page 11: ...Wash for cassettes hinged instruments or heavily soiled instruments Cycle screen Cycle name appears at top of screen Progress bar shows progress of the cycle as percentage of remaining time vs remain...

Page 12: ...______ Cycle aborted acknowledge screen If cycle aborted due to cycle fault or door opening display will indicate problem Press the OK button to return to the main menu screen ________________________...

Page 13: ...1 13 Overview of Menus SciCan HYDRIM SH11MR311 User Page 15 Technician Page 21 Setup Page 17 Error History Cycle Count Unit No Language Selection Set Regeneration Set Button Beep Screen Saver Temperat...

Page 14: ...Drying Time The drying time for P2 and P3 can be set independently Touch the cycle for which you want to set the drying time Use the up and down arrows to select the desired drying time Choose from 1...

Page 15: ...llows the user to see how many cycles have been run on the Hydrim Touch cycle count and the number of cycles completed and aborted will be displayed This cannot be reset Touch the back arrow to return...

Page 16: ...rrows to move through the choices touch the desired option to select it For more detail on each Setup Menu option refer to the following sections Unit Number If you have more than one Hydrim in your o...

Page 17: ...change the value The default setting is 0 See installation instructions for further information Touch the back arrow to return to the previous menu ___________________________________________________...

Page 18: ...in degrees Celsius or degrees Fahrenheit Touch the up and down arrows to select the desired choice The default is degrees Celsius Touch the back arrow to return to the previous menu __________________...

Page 19: ...r errors Test Cycle An abbreviated cycle that uses all pumps and valves Cannot be used to process instruments Component Test Allows individual testing of the following components Circulation Pump ON O...

Page 20: ...ols Cycle Settings Offers a submenu of eight settings Set Drying Time P2 Wash 10 min choose from 1 to 20 min P3 Heavy Duty Wash 10 min choose from 1 to 20 min Set Regeneration Sets water softener sett...

Page 21: ...se from 45 to 80 C P3 Heavy Duty Wash 50 C choose from 45 to 80 C Last Rinse Temperature P2 Wash 60 C choose from 30 to 93 C P3 Heavy Duty Wash 60 C choose from 30 to 93 C Repeater Mode In the ON posi...

Page 22: ...nu touch 7919 and press EN Password is shown as characters Touch CL to clear the last number entered Press the back arrow to return to the previous menu The Technician Menu consists of these six choic...

Page 23: ...evious menu Error History screen allows the user to see the last four errors Press the down arrow to move to the next page error Press the X in the lower left hand corner to clear the error history Th...

Page 24: ...e Y7 ON OFF Cold W Valve Y1 ON OFF Rinse Aid Valve ON OFF Waste Pump M2 ON OFF Up arrow pressed next device Down arrow pressed previous device On button pressed set device On Off button pressed set de...

Page 25: ...Dryer Motor OFF Dryer Heater OFF Drain Pump ON Dosing OFF Salt V OFF Time remaining in cycle Inputs Chamber Temperature is 27 C Chamber Full Switch is OFF Detergent Switch is OFF Chamber Overflow Swi...

Page 26: ...and P3 can be set independently Touch the cycle for which you want to set the drying time Use the up and down arrows to select the desired drying time Choose from 1 to 20 min Press default if you wish...

Page 27: ...select the desired dosing setting for the cycle selected Do Not change default setting unless directed by SciCan Press default to return to factory default Press the back arrow to return to the Techni...

Page 28: ...time for the cycle selected Do not change default setting unless directed by SciCan Press default to return to factory default Press the back arrow to return to the Technician Menu P2 P3 rinse times...

Page 29: ...t change default setting unless directed by SciCan Press default to return to factory default Press the back arrow to return to the Technician Menu P P2 P3 last rinse temperature Min Value 30 C Max Va...

Page 30: ...elect 5849 and press EN Touch CL to clear the last number entered Press the back arrow to return to the previous menu __________________________________________________________________________________...

Page 31: ...m PRES ON Chm OVER OFF Door CLSD Salt SW No Salt ColdV OFF HotV ON CircPUMP ON ChmHEATER OFF DryerMOTOR OFF DryerHEATER OFF DrainPUMP OFF Dosing OFF SaltV OFF Time remaining in cycle Outputs Cold wate...

Page 32: ...in the first 6 minutes of filling circulation pump not running yet if hot water missing or at low pressure unit will switch to cold water after 4 minutes if chamber full switch is not activated Press...

Page 33: ...d or software failure CF11 No Detergent Flow 1 Detection The flow switch did not detect detergent flowing CF11 will not be displayed if the flow switch doesn t detect detergent during the rinse phase...

Page 34: ...ve not completed the cycle should not be used and should be reprocessed Cover Removal Top Cover Open front door and remove two screws under front of top cover Lift up top cover to remove Lower Front C...

Page 35: ...g Reinstall new Flow Switch in reverse order Flow Switch Part 17137 02 Dosing Pump The Dosing Pump is located behind the cover below the front door Remove screw from top center of cover and locate the...

Page 36: ...ind the cover below the front door remove the sensor from the chamber and unplug the red wiring Chamber Temp Sensor Part 22816 12 Door Latch Remove top cover Disconnect Door Latch in front of unit and...

Page 37: ...needed IO Board Part 01 109783S Logic Board Part 01 110217S Power Supply Board Remove top cover and right side panel to access Power Supply Board Power Supply Board Part 01 109782S Dryer Motor and Hea...

Page 38: ...Remove the top and rear covers to access the Chamber Pressure Switch Chamber Pressure Switch Part 22859 10 Cold Hot Water Inlet Valves Remove the top and rear covers to access the Cold Hot Water Inlet...

Page 39: ...1 39 Drain Pump Remove top cover and right side panel to access the Drain Pump Drain Pump Part 21942 00 Circulation Pump Remove top cover rear and left side panels to access the Circulation Pump Circu...