![SciCan HYDRIM L110w Service Manual Download Page 35](http://html1.mh-extra.com/html/scican/hydrim-l110w/hydrim-l110w_service-manual_1217173035.webp)
Hydrim L110w Service Guide 95-109977-1
…
35
Component Replacement
Flow Switch
The Flow Switch is located behind the solution box. Disconnect the
solution box from the black cap and hose assembly and remove box
from solution door. Remove screw from back panel inside solution
door. Lift up on mounting bracket and turn so Flow Switch is
accessible. Cut tie wrap holding Flow Switch to mounting bracket and
remove tubing’s using Hemostats to pinch tubing’s so they don’t leak.
Gently pull Flow Switch towards you until electrical plug is accessible
then disconnect plug. Reinstall new Flow Switch in reverse order.
Flow Switch Part # 17137.02
Dosing Pump
The Dosing Pump is located behind the cover below the front
door. Remove screw from top center of cover and locate the
Dosing Pump on the right side. Remove two mount screws to
access the Dosing Pump.
Dosing Pump Part # 21972.01
Water Heater
The Water Heater connections are located behind the cover below the front door. The heater is located in
the chamber and locked in place with the double U plate from outside. After removing the mounting nuts
and U plate the Water Heater is removed from inside the chamber underneath the filter screen. Remove
safety switch probe before removing the heating element.
Water Heater Part # 23012.01
U Plate
Safety Switch Probe
Mounting Plate & Gasket