Hydrim L110w Service Guide 95-109977-1
…
25
The Debug cycle should be used when troubleshooting to view the IO status of components.
P2 Wash 24 min
Draining
Chm T 27ºC
Chm Full=OFF Det SW=OFF
Chm Over=OFF Chm Pres=OFF
Door=CLSD Salt SW=ON
Cold V=OFF Hot V=OFF
Circ Pump=OFF
Chm Heater=OFF
Dryer Motor=OFF
Dryer Heater=OFF
Drain Pump=ON Dosing=OFF
Salt V=OFF
Time remaining in cycle
Inputs
Chamber Temperature is 27ºC
Chamber Full Switch is OFF
Detergent Switch is OFF
Chamber Overflow Switch is OFF
Chamber Pressure Switch is OFF
Door latch is CLOSED
Salt Switch is OFF
Outputs
Cold water valve is OFF (closed)
Hot water valve is OFF (closed)
Circulation pump is OFF
Chamber heater is OFF
Dryer motor is OFF
Dryer heater is OFF
Drain pump is ON
Dosing is OFF
Salt valve is OFF