![SciCan HYDRIM L110w Service Manual Download Page 3](http://html1.mh-extra.com/html/scican/hydrim-l110w/hydrim-l110w_service-manual_1217173003.webp)
Hydrim L110w Service Guide 95-109977-1
…
3
Cycle Description
P1
Rinse & Hold Cycle*
P2
Regular Cycle
P3
Heavy Duty Cycle
Description
Use to keep
instruments moist when
not being washed for 1
hour
Use for moderately
soiled loose
instruments
Use for heavily soiled
instruments and
cassettes
Cold Prewash
<45ºC / 113ºF
N/A
<45ºC / 113ºF
Wash
N/A
50ºC / 122ºF
5 minutes
50ºC / 122ºF
9 minutes
Final Rinse
60ºC / 140ºF
60ºC / 140F
60ºC / 140ºF
Dry
N/A
1 to 20 min.
Default 10 min.
1 to 20 min.
Default 10 min.
Total Time**
without drying
9 minutes
16 minutes
25 minutes
Water Consumption***
11.5L / 3 gallons
19L / 4.75 gallons
25L / 6.25 gallons
*
This is not a wash cycle. Always run a wash cycle following a rinse and hold cycle.
**
Cycles times depend on the temperature of incoming water.
***
Water consumption depends on the regeneration Level
Specifications
Height, freestanding: 850 mm / 33.5"
Height, built-in: 830 mm / 32.75"
Width: 600 mm / 23.75"
Depth: 600 mm / 23.75"
Depth with door open: 1200 mm / 47"
Weight: 80 kg / 176 lbs.
Running Noise: 60 dB(A)
Hot and cold water connections: 3/4”
Inlet water pressure: 1-10 bar/14.5-145 psi
Incoming hot water temperature: 50-70°C / 122-158°F
Drain: 3/4”
Water softener: 1 kg / 2.2 lbs salt capacity
Drying System: Heater 1 kW
Electrical Connection: 208-240 V 60 Hz 15 A
Equipment pollution degree: Pollution Degree 2
Equipment installation category: Installation category II
Maximum relative humidity: 80% for temp up to 31°C / 88°F
50% for temp up to 40°C / 104°F
Operating temperature range: -5°C to 40°C / 23-104°F
Max. altitude: 2000m / 6561.7 ft.
Mains supply: + / -10% of nominal