Hydrim L110w Service Guide 95-109977-1
…
20
View IO Status: Shows the status of each component in the unit.
Chamber Full Switch – ON/OFF
Salt Switch – OK/No Salt
Rinse Aid – ON/OFF (not used)
Chemical Sensor – ON/OFF
Chamber Overflow – ON/OFF
Chamber Pressure – ON/OFF
Door Switch – ON/OFF
Chamber Temperature Celsius
CTS (Clear to Send) – ON/OFF
Set debug screen: Shows IO status when cycle is running. (see troubleshooting tools)
Cycle Settings:
Offers a submenu of eight settings.
Set Drying Time:
P2 Wash
10 min ↑↓ (choose from 1 to 20 min)
P3 Heavy Duty Wash
10 min ↑↓ (choose from 1 to 20 min)
Set Regeneration: Sets water softener setting.
0 ↑↓ (choose from 0 to 7)
Chemical Setup:
Prewash
5 sec. ↑↓ (choose from 0 to 15 sec.)
Wash
14 sec. (choose from 0 to 30 sec.)
High Temp. Wash
15 sec. (choose from 0 to 30 sec.)
Rinse
10 sec (choose from 0 to 15 sec.)
Set Wash Time:
P2 Wash
5 min ↑↓ (choose from 5 to 15 min)
P3 Heavy Duty Wash
9 min ↑↓ (choose from 5 to 15 min)
Set Rinse Time:
P2 Wash
1 min ↑↓ (choose from 1 to 10 min)
P3 Heavy Duty Wash
1 min ↑↓ (choose from 1 to 10 min)