OPERATING MANUAL STIMAWELL® 120MTRS EN
41
When you mark a patient and touch on
OK
, you select the patient for therapy (please see
Section 9.3).
With
Cancel
you return to Patient Administration without having implemented the changes.
Please Note:
If you wish to change the mat size, with selected patient data set touch in the
main menu on the display
Representation Of The StimaWELL
®
Stimulation Mat
. Change the
mat size and confirm with
OK
. Finally, touch on
Save
, to accept permanently the changes for
the patient. If the mat size has been changed from M to L or from S to M or L, it is recom-
mended to carry out a renewed calibration. Otherwise the proportional intensity of the two
lowest channels will correspond with the standard setting levels. Please bear in mind here that
you cannot change the mat size if a patient has been selected in the mode
Automatic Program
Selection Based On Indications
with the therapeutic measures
Pain Therapy
.
9.4.1
History Evaluation
After you have touched on
History
, all com-
pleted therapies of the patient are displayed
in a list. Each entry consists of the date of the
therapy, the therapy measures including the
flow form and localisation as well as the num-
ber of the therapy sessions. Please note that
only fully completed sessions are included in
the therapy history and not those which were
for any reason discontinued before the ses-
sion was completed.
To select a therapy, mark by touching. Activate
Delete
, to delete the selected therapy from
the list. If you touch on
Evaluation
, the detailed data for the selected therapy is displayed.
In the upper part of the screen you see the
number of the therapy sessions together with
the respective dates. Below is displayed how
frequently the patient, subsequent to the
therapy session, has indicated whether
his/her pain level was better, the same or
worse.
Should Pain Therapy be involved in the thera-
py history, further details can be found in the
evaluation.