26
OPERATING MANUAL STIMAWELL® 120MTRS EN
will be cancelled by touching on the button
Delete
. With
OK
the input is confirmed and
with
Cancel
you exit the input field without having implemented the changes.
To input a size or a weight please touch on the arrow fields left and right of the respective
field. The standard data are 180 cm / 80 kg and can be changed to values in the range 120 –
250 cm / 30 – 200 kg.
To input further data please touch on the button
Extended/Basis
. With this button you
can scroll through several pages of a window.
Simply touch on the field you wish to fill out in
order to open the respective input window.
Once you have entered all the required data
you arrive with
Continue
in the window for
determining the mat size. With
Back
you re-
turn to
Patient Administration
without having
implemented the changes.
9.2.2
Setting The Mat Size
After confirming the patient data there opens
automatically the window for the setting of
the mat size (Please also see Section 6.3).
According to the size of your patient select the
mat size S, M or L. Confirm your selection with
Continue
or return with
Back
to the window
for the input of patient data.
9.2.3
Calibration / Setting The Partial Intensity For The Channels
The calibration follows after selecting the mat
size. If you have gone to the
Therapy Options
and activated the
Skip Calibration
selection
box, a window will open allowing you to select
Body Side Related Calibration
or
Skip Calibra-
tion
. The calibration is thereby a fine tuning of
the intensity of the individual channels. It is for
this reason that we would advise against skip-
ping the calibration. If you should nevertheless
touch on
Skip Calibration
, you arrive directly in
the window
Select Therapy Mode
(please see Section 9.2.4).