OPERATING MANUAL STIMAWELL® 120MTRS EN
17
11
Button
Pause
Pause the running therapy
12
Button
Stop
Stop the running therapy
13
Display
Representation Of The StimaWELL
®
Stimulation Mat
Display of the active channels
14
Display
Selected Therapy Program
15
Button
Favourites Star
Include a program in the favourites list / delete from favourites list
16
Display
Patient’s Name
17
Display
Time
Setting The Mat Temperature
6.1
You can set the temperature for the StimaWELL
®
Stimulation Mat in the range from 20 – 40 °C
in steps of 5°C. For this, touch on the display
Temperature
. On the lower edge of the display
will appear the [+] - [-] symbols with which the temperature can be changed. Now touch the
[+] or [-] to increase/decrease the temperature and/or to switch off the mat heating com-
pletely. To change the values more quickly, maintain pressure on the respective button. The
temperature will then continue to be decreased/increased until the finger is removed from
the button or the limit value has been attained.
If you touch the field at a temperature of 20°C on the [-] sym-
bol, the mat heating is switched off. Renewed touching on the
[+] symbol will reactivate the heating. If you touch the field at
a temperature of 20°C on the [-] symbol, the mat heating is
switched off. Renewed touching on the [+] symbol will reacti-
vate the heating.
Important Note
:
Depending on the ambient temperature,
the mat heating needs approximately 20 minutes to heat the
StimaWELL
®
Stimulation Mat from 20°C to 40°C. Therefore switch on the heating well before
your patient’s appointment time.
When switching on the control unit, the mat heating is nor-
mally inactive. If you select the
Technical Options
menu, the
option
Activate Heating When Switching ON
(please see Sec-
tion 12.2.2), the mat heating will be set and then activated
when switching on the StimaWELL
®
120MTRS Control Unit to
40°C.