All rights reserved! Non authorized copy and/or not agreed usage of document prohibited.
Page 31
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2,5 GHz
[V1] 3 V
[E1] 3 V/m
Portable and mobile RF communications equipment
should be used no closer to any part of the device,
including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance:
d=
d=
80 MHz to 800 MHz
d=
800 MHz to 2,5 GHz
where P is maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Recommended separation distance is 0,04m (4cm).
LED blinking on charging dock during Proximity field
immunity test.
At 710 MHz changes on graph appeared, without
influence on time-base of recording.
Field strength from fixed RF transmitters as determined
by an electromagnetic site survey, should be less than the
compliance level in each frequency range.
Interference may occur in the vicinity of equipment
marked with the following symbol:
P
V
]
5
,
3
[
1
P
E
]
5
,
3
[
1
P
E
]
7
[
1