Page 28
All rights reserved! Non authorized copy and/or not agreed usage of document prohibited.
NOTE: When the Savvy sensor is not in use, it is necessary to be placed in the charging dock and
in charging mode in order to ensure that the battery will not be discharged. Discharged battery
void Guarantee.
Figure 20: Savvy sensor placed in the charging dock connected to a Charger, while charging.
9.2.
Visual inspection
Visually inspect the device and cable assemblies for the following:
•
Savvy sensor casing not broken or cracked
•
Charging dock casing not broken or cracked
•
Charger casing not broken or cracked
•
Charger cable sheathing and connectors undamaged. No kinks in the cable.
Defective units or damaged cables void guarantee and must be replaced immediately prior
further use.
9.3.
Cleaning instructions
The Savvy sensor, Charging dock, Charger can be cleaned and disinfected only when they are
disconnected from power supply and only when fully dry, they can be reconnected to power
supply.
The Savvy sensor can be cleaned and disinfected with medical ethyl alcohol or special cleaning
agent for medical devices, to ensure compliance with the following requirements:
•
Biocompatibility of the housing,
•
Aging of the polypropylene,
•
Compatibility with adhesive endurance.
The damage caused when using the wrong cleaning agents void product/producer warranty.
It is allowed to wipe the device with wet cloth. While the Savvy sensor is water resistant, do not
immerse it in water.
9.4.
MobECG application - User maintenance
User must keep the Android environment as specified in the requirements:
•
Adequate storage space is maintained on the PDA and personal computer;
•
Internet connection on the PDA and the PC is established as required – when
transferring measurements to PC or CS;
•
Bluetooth radio is enabled on the PDA.
•
The PDA is turned on while recording, while the screen can be off.