10
The time set here applies to alarms 1, 2 and 3.
a7 (A7)
Wind direction alarm criteria
Measurement item to determine the wind direction alarm
conditions
0: No alarm output
1: Determined by the instantaneous
wind direction
2: Determined by the mean wind
direction
a8 (A8)
Wind direction alarm section 1, start point setting
value
Starting angle for the wind direction alarm section 1
Setting range: 0 to 359(°)
a9 (A9)
Wind direction alarm section 1, end point setting
value
Ending angle for the wind direction alarm section 1
Setting range: 0 to 359(°)
a0 (A0)
Wind direction alarm section 2, start point setting
value
Starting angle for the wind direction alarm section 2
Setting range: 0 to 359(°)
b1
Wind direction alarm section 2, end point setting
value
Ending angle for the wind direction alarm section 2
Setting range: 0 to 359(°)
b2
Setting value for the wind direction alarm
confirmation time
The alarm is activated if the wind direction readings
remain within the angle range set by section 1 or 2 for
this time period.
Setting range: 0 to 60 (sec)
b3
Setting value for the wind direction alarm
cancellation time
After activation, the alarm is cancelled if the wind
direction readings remain outside the angle range set by
section 1 or 2 for this time period.
Setting range: 3 to 60 (sec)
b4
Display item in the upper wind speed display section
3: Instantaneous wind speed, fixed
b5
Display item in the lower wind speed display section
4: Mean wind speed, fixed
b6
Display item in the wind direction display section
1: Instantaneous wind direction
2: Mean wind direction
III-6-3 Setting mode
Setting values are modified by operating the buttons.
Setting mode display
Wind Speed & Direction Indicator with Alarm
Mean
Mean
Instant
Wind Direction
Instant
Mode code is changed with
V buttons
Setting value is changed with
V buttons.
BUZZER [ENT] button is use in fix.
Alarm