10
Important Note:
If the deviation is too high, the analyzer is not
adjusted. In this case, please inform your local
Sartorius office, dealer or service center.
§
Open the sample chamber by pressing the
k
key
>
Allow the moisture analyzer to cool down.
§
Snap the disk holder into the adjustment disk and
remove the disk. Caution: Equipment is extremely
hot after drying!
To avoid severe burns, do not touch until
sufficiently cooled.
!
Allow the adjustment disk to cool on a clean,
fire-proof pad.
--------------------
27.05.2009
10:23
Mod.
MA100C000230V
Ser. no.
90805350
Ver. no.
01-38-02
ID
--------------------
2-PT.ADJ.:
--------------------
Ref ID
90801234
L.temp.
80'C:
78'C
U.temp. 140'C: 163'C
Adjustment
completed
--------------------
27.05.2009
10:23
Name:
--------------------
Summary of Contents for YTM05MA
Page 16: ...16 Sartorius Factory Calibration Certificate Example...
Page 30: ...30 Werkskalibrierzertifikat Beispiel...
Page 44: ...44 Certificat de calibrage d usine Exemple...
Page 58: ...58 Certificado de calibraci n de f brica Ejemplo...
Page 72: ...72 Certificato di calibrazione di fabbrica Esempio...
Page 73: ...73...