8
§
Use the disk holder to remove the adjustment
disk from its carrying case.
§
Insert the adjustment disk under slight pressure.
The marking should be facing towards the user.
§
Close the sample chamber by pressing the
k
key.
§
Press the
T
key; the following is displayed:
“Heater adjustment”.
§
Start “Heater adjustment” by pressing the
O
soft key.
§
Select “2-point temperature adjustment”; start by
pressing the
O
soft key.
§
If required, enter the serial number (ref. ID number)
(4). Confirm by pressing the soft key [
Ref ID
].
The reference ID number is a serial number;
for example: 11901617 (on the bottom side of the
temperature adjustment disk).
SETUP
isoTEST
Heater adjustment
SETUP
isoTEST
Heater adjustment
SETUP
isoTEST
HEATER
2-point temperature adjustment
1-point temperature adjustment