
15
Error Codes
“Error 53”
The temperature adjustment disk has not been inserted properly.
“L”
Star-shaped retainer missing.
The temperature adjustment routine has not been selected yet.
Press the
T
key.
“Canceled”
The sample chamber was opened during the measurement.
“Outliers”
The temperature adjustment disk is contaminated.
The moisture analyzer is not ready for operation.
The moisture analyzer has been running too long (overheated).
The moisture analyzer is defective; please inform your local
Sartorius office, dealer or service center.
“too many charac.”
You have entered more than 14 characters for the reference
ID no.
“no entry”
You have not entered a temperature for the 1-point adjustment.
“too high”
The temperature entered is too high.
“too low”
The temperature entered is too low.
Summary of Contents for YTM05MA
Page 16: ...16 Sartorius Factory Calibration Certificate Example...
Page 30: ...30 Werkskalibrierzertifikat Beispiel...
Page 44: ...44 Certificat de calibrage d usine Exemple...
Page 58: ...58 Certificado de calibraci n de f brica Ejemplo...
Page 72: ...72 Certificato di calibrazione di fabbrica Esempio...
Page 73: ...73...