12
§
Use the disk holder to remove the adjustment
disk from its carrying case.
§
Insert the adjustment disk under slight pressure.
The marking should be facing towards the user.
§
Close the sample chamber.
§
Press the
u
key; select “isoTEST” SETUP by
pressing the
x y
keys. Confirm by pressing
the
u
key. The following appears on the display
“Heater adjustment”. Press the
u
key again.
§
Select “2-point temperature adjustment”; confirm
by pressing the
u
key.
§
Press the
u
key again to activate the temperature
adjustment routine.
> The heating symbol appears on the display; the
heating time and the previous drying time are
displayed continuously.
SETUP
isoTEST
Heater adjustment
SETUP
isoTEST
HEATER
2-point temperature adjustment
1-point temperature adjustment