Cubis MSA User Manual
181
Specifications
Model-specific Data
Verified models with EC Type Approval Certificate: Precision balances
Models
MSA6202S-
MSA6202P-
MSA5202S-
MSA4202S-
0CE
0CE
0CE
0CE
Accuracy class*
K K K K
For verified models: EC Type-Approval Certificate D09-09-015, Type: MSX
Scale interval d*
g
0.01
0.01/0.02/0.05
0.01
0.01
Weighing capacity max*
g
6200
1500/3000/6200
5200
4200
Calibration value e*
g
0.1
0.1
0.1
0.1
Min. load min*
g
0.5
0.5
0.5
0.5
Tare equalization range (subtractive)
100% from max. weighing capacity
Application range according to DIR*
g
0.5 – 6200
0.5 – 6200
0.5 – 5200
0.5 – 4200
Min. initial weight**
g
12
12
10
12
Typical stabilization time
s
1
1
0.8
0.8
Typical measurement time
s
1.5
1.5
1
1
Application range (temperature):
With “isoCAL” function
+5 to +40 °C
+5 to +40 °C
+5 to +40 °C
Without “isoCAL” function
+10 to +30 °C
+10 to +30 °C
+10 to +30 °C
Display result
(depending on the set filter level)
0.1 – 0.4
Weighing pan size (W
D)
mm
206
206
206
206
140
140
206
206
Protection
Protected against dust and water
Models
MSA2202S-
MSA1202S-
MSA12201S-
MSA8201S-
MSA5201S-
0CE
0CE
0CE
0CE
0CE
Accuracy class*
K K
K
K K
For verified models: EC Type-Approval Certificate D09-09-015, Type: MSX
Scale interval d*
mg
10
10
100
100
100
Weighing capacity max*
g
2200
1200
12,200
8200
5200
Calibration value e*
g
0.1
0.1
1
1
1
Min. load min*
g
0.5
0.5
5
5
5
Tare equalization range (subtractive)
100% from max. weighing capacity
Application range according to DIR*
g
0.5 – 2200
0.5 – 1200
5 – 12,200
5 – 8200
5 – 5200
Min. initial weight**
g
12
12
100
100
100
Typical stabilization time
s
0.8
0.8
0.8
0.8
0.8
Typical measurement time
s
1
1
1
1
1
Application range (temperature)
With “isoCAL” function: +5 to +40 °C | Without “isoCAL” function: +10 to +30 °C
Display result
(depending on the set filter level)
0.1 – 0.4
Weighing pan dimensions (W + D)
mm
206
206
Protection
Protected against dust and water
* DIR = Directive 90/384/EEC on non-automatic weighing instruments used within the European Economic Area
**
= Typical min. initial weighing according to USP (Unites States Pharmacopeia), USP31-NF26
Summary of Contents for MSA3.6P
Page 192: ...192 Cubis MSA User Manual Declarations of Conformity...
Page 193: ...Cubis MSA User Manual 193 Declarations of Conformity...
Page 194: ...194 Cubis MSA User Manual Declarations of Conformity...
Page 195: ...Cubis MSA User Manual 195 Declarations of Conformity...
Page 196: ...196 Cubis MSA User Manual Declarations of Conformity...
Page 197: ...Cubis MSA User Manual 197 Plates and Markings...
Page 198: ...198 Cubis MSA User Manual Plates and Markings...
Page 199: ...Cubis MSA User Manual 199 Plates and Markings...
Page 200: ...200 Cubis MSA User Manual Plates and Markings...