178
Cubis MSA User Manual
Specifications
Model-specific Data
Verified models with EC Type Approval Certificate: Semi-microbalances 0.01 mg
Model
MSA225S-0CE
MSA225P-0CE
MSA125P-0CE
Accuracy class*
k k k
For verified models: EC Type-Approval Certificate D09-09-015, Type: MSX
Scale interval d*
mg
0.01
0.01/0.02/0.05
0.01/0.1
Weighing capacity max*
g
220
60/120/220
60/120
Calibration value e*
mg
1
1
1
Min. load min*
mg
1
1
1
Tare equalization range (subtractive)
100% from max. weighing capacity
Application range according to DIR*
g
0.001 – 220
0.001 – 220
0.001 – 120
Min. initial weight**
mg
20
20
20
Typical stabilization time
s
2
2
2
Typical measurement time
s
6
6
6
External standard calibration value
g
200 (E2)
200 (E2)
100 (E2)
(min. accuracy class)
Application range (temperature)
With “isoCAL” function: +5 to +40 °C | Without “isoCAL” function: +15 to +25 °C
Adaptation to ambient conditions
By selection of 1 of 4 optimized filter levels
Display result (depending on the set filter level)
0.2 – 0.4
Weighing pan size (W
D)
mm
85
85
Weighing chamber height (draft shield DU) mm
261
Protection
Protected against dust and water
Verified models with EC Type Approval Certificate: Analytical balances 0.1 mg
Model
MSA524S-
MSA524P-
MSA324S-
MSA224S-
MSA324P-
MSA124S-
0CE
0CE
0CE
0CE
0CE
0CE
Accuracy class*
k k k k k k
For verified models: EC Type-Approval Certificate D09-09-015, Type: MSX
Scale interval d*
mg
0.1
0.1/0.2/0.5
0.1
0.1
0.1/0.2/0.5 0,1
Weighing capacity max*
g
520
120/240/520 320
220
80/160/320 120
Calibration value e*
mg
1
1
1
1
1
1
Min. load min*
mg
10
10
10
10
10
10
Tare equalization range (subtractive)
100% from max. weighing capacity
Application range according to DIR*
g
0.01–520
0.01–520
0.01–320
0.01–220
0.01–320
0.01–120
Min. initial weight**
mg
120
120
120
120
120
120
Typical stabilization time
s
1
1
1
1
1
1
Typical measurement time
s
3
3
3
3
3
3
External standard calibration value
g
500
500
200+100
200
200 +100
100
(min. accuracy class)
(E2
(E2)
(E2)
(E2)
(E2)
(E2
Application range (temperature)
With “isoCAL” function: +5 to +40 °C | Without “isoCAL” function: +15 to +25 °C
Display result (depending on the set filter level)
0.1 – 0.4
Weighing pan size (W
D)
mm
85
85
Weighing chamber height (draft shield DU) mm
261
Protection
Protected against dust and water
* DIR = Directive 90/384/EEC on non-automatic weighing instruments used within the European Economic Area
**
= Typical min. initial weighing according to USP (Unites States Pharmacopeia), USP31-NF26
Summary of Contents for MSA3.6P
Page 192: ...192 Cubis MSA User Manual Declarations of Conformity...
Page 193: ...Cubis MSA User Manual 193 Declarations of Conformity...
Page 194: ...194 Cubis MSA User Manual Declarations of Conformity...
Page 195: ...Cubis MSA User Manual 195 Declarations of Conformity...
Page 196: ...196 Cubis MSA User Manual Declarations of Conformity...
Page 197: ...Cubis MSA User Manual 197 Plates and Markings...
Page 198: ...198 Cubis MSA User Manual Plates and Markings...
Page 199: ...Cubis MSA User Manual 199 Plates and Markings...
Page 200: ...200 Cubis MSA User Manual Plates and Markings...