Task Management
Task Management
In the Task management menu (TASK), you can:
– Start a task
– Configure a new task
– Edit an existing task (modify, copy, delete)
For
task
configuration, there are numerous
applications
available, e. g. weighing,
parts counting, calculation. Each task contains at least one of these applications.
However, it can also have a combination of several applications. Each applica-
tion can be used either with their factory settings or you can also modify them as
desired. The
Weighing
application is available as a basic function for every task.
Each task also contains the individual balance settings and the individually config-
ured printout.
The options available to you in Task Management depend upon your user access
rights: An
Administrator
can use all options while a
User
can only use limited
options (see “User Management” in the User Management chapter). If no users have
been defined, everything is accessible, there are no rights.
The administrator can
create tasks with restricted use
, i.e. tasks that can only be
used by certain users or only by the administrator himself.
Global tasks
can be used by all users.
Separate tasks
are only accessible to the users who created them.
Exception: A user can copy global tasks to use them afterwards as a local task.
Option
Administrator User
Configure, modify, delete separate tasks
i
Configure, modify, delete global tasks
i
Execute or copy separate tasks
i
Execute or copy global tasks
i
i
D
t
To enter the Task Management, press the key
6#5-
D
.
y
A list of available tasks is displayed.
Initially, this overview is empty. Once tasks have been configured, they will be
displayed here.
t
When you touch
5VCTV
, the task currently selected will be started (indentifiable
by the dark background).
Importing tasks:
You can also load preconfigured tasks to the balance:
1)
Unpack the desired ZIP archive file to the root directory (Root Dir)
of a SD card after download from the Sartorius homepage or from
the included CD.
2)
In the Balance menu, submenu “Import/Export data”:
Execute “Import files from the SD card.”
3) Insert the SD card into the balance.
4)
Select the corresponding import file and active the task download via
0GZV
. Then exit the menu.
Summary of Contents for MSA3.6P
Page 192: ...192 Cubis MSA User Manual Declarations of Conformity...
Page 193: ...Cubis MSA User Manual 193 Declarations of Conformity...
Page 194: ...194 Cubis MSA User Manual Declarations of Conformity...
Page 195: ...Cubis MSA User Manual 195 Declarations of Conformity...
Page 196: ...196 Cubis MSA User Manual Declarations of Conformity...
Page 197: ...Cubis MSA User Manual 197 Plates and Markings...
Page 198: ...198 Cubis MSA User Manual Plates and Markings...
Page 199: ...Cubis MSA User Manual 199 Plates and Markings...
Page 200: ...200 Cubis MSA User Manual Plates and Markings...