77
Operating the Scale
Define the Calibration Weight:
FCA Models
Step
Press key(s)
Display/Output
(or follow instructions)
1. Select Setup menu
M
2. Select “Balance/scale functions”
O
soft key
3. Select “Calibration/adjustment”
O
soft key
4. Select parameter for external
q
soft key 5 x
weight
O
soft key
5. Select “Cal./adj. wt.”
q
soft key
3000 = last setting selected
6. Enter calibration weight
5 0 0 0
(e.g., 5000 g)
and save
7. Save the calibration weight
l
soft key
8. Exit the Setup menu
oo
soft key