173
Error Codes
Error Codes
Error codes are displayed in the main display or text line for 2 seconds.
The program then returns automatically to the previous status.
Display
Cause
Solution
No segments appear on the display
No AC power is available
Check the AC power supply
The AC adapter is not plugged in
Plug in the AC adapter
Automatic shutoff configured in
Press
e
to switch on the scale
Setup (code
8 7 1
)
or select code
8 7 2
in Setup (“no automatic shutoff”)
H
The load exceeds the
Unload the scale
scale capacity
L
or
Err 54
The weighing pan is not in place
Place the weighing pan
on the scale/may have to be turned
off and on again if “
L
” is displayed
Err 01
Data output not compatible with
Change the configuration in
O
Display range
output format
the Setup menu
Err 02
Calibration/adjustment condition
Calibrate only when zero
Cal. n. possible
not met, e.g.,
is displayed
– The scale was not tared
Press
t
to tare
– The scale is loaded
Unload the scale
Err 03
Calibration/adjustment could not
Allow the scale to warm up
Cal.
/
adj. interrupt
be completed within a certain time
again and repeat the
adjustment process
Err 06
Built-in calibration weight
Contact your local Sartorius
Int. wt. defective
is defective
Service Center
Err 07
Function not allowed in scales
Contact your local Sartorius
Function blocked
verified for use in legal metrology
Service Center for information
on having the settings changed
Err 08
*
The load on the scale is too
Check whether the “power-on
oO
zero range
heavy to zero the readout
zero range” is set
Err 09
*
Taring is not possible when the gross
Zero the scale
o
0 not allowed
weight is
≤
zero
Err 10
Tare key and 2nd tare memory are
Press
c
to clear the formulation
Tare fct. blocked
blocked when there is data in the
application; the tare key and 2nd
tare memory for the formulation
tare memory are then accessible
application
Err 11
Tare memory not allowed
Check the tare value entered
Tare2 blocked
Err 12
Tare memory greater than weighing
Check sample/container
Tare2
O
Max.
range or range limits
Err 17
Internal adjustment is not possible
Reduce the preload or
Adj.-wt.
O
Max.
because preload is too heavy
change the configuration
Err 30
Interface port for printer output
Contact your local
Print fct. blocked
is blocked
Sartorius Service Center
* = occurs only via the SBI interface (ESC f3_/f4_)
Summary of Contents for Factory Series
Page 172: ...172 ...
Page 193: ...193 Overview ...
Page 194: ...194 ...
Page 195: ...195 Overview ...
Page 196: ...196 ...
Page 197: ...197 Overview ...
Page 198: ...198 ...
Page 199: ...199 Overview ...
Page 200: ...200 ...
Page 201: ...201 Overview ...
Page 202: ...202 ...
Page 203: ...203 Overview ...
Page 204: ...204 ...
Page 205: ...205 Overview ...
Page 206: ...206 ...
Page 207: ...207 Overview ...
Page 208: ...208 ...
Page 209: ...209 Overview ...
Page 210: ...210 ...