Line Format (Data Output Format)
You can output the values displayed
in the measured value line and
the weight unit with or without a
data ID code
Example: Without data ID code
+
253 pcs
Example: With data ID code
Qnt
+
253 pcs
Configure this parameter in the
Setup menu (Setup: Basic settings:
Printout configuration: Line format).
The output with data ID code has
16 characters; without data ID code,
22 characters.
Output Format With 16 Characters
Display segments that are not activated are output as spaces. Characters
without a decimal point are output without a decimal point.
The following characters can be output, depending on the characters
displayed on the scale:
Normal Operation
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
+ * D D D D D D D D * U U U CR LF
or
–
.
.
.
.
.
.
.
.
* * *
or
*
* * * * * * * *
*:
Space
D:
Digit or letter
U:
Unit symbol
CR:
Carriage return
LF:
Line feed
Special Codes
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
* * * * * *
–
–
* * * * * * CR LF
or
H H
or
L
L
or
C
*:
Space
– –:
Weight
H:
Overload
H H:
Overload in checkweighing
L:
Underload
L L:
Underload in checkweighing
C:
Calibration/adjustment
Error Codes
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
* * *
E
r
r
* */# # # * * * * CR LF
*:
Space
# # #: Error code number
Summary of Contents for Factory Series
Page 172: ...172 ...
Page 193: ...193 Overview ...
Page 194: ...194 ...
Page 195: ...195 Overview ...
Page 196: ...196 ...
Page 197: ...197 Overview ...
Page 198: ...198 ...
Page 199: ...199 Overview ...
Page 200: ...200 ...
Page 201: ...201 Overview ...
Page 202: ...202 ...
Page 203: ...203 Overview ...
Page 204: ...204 ...
Page 205: ...205 Overview ...
Page 206: ...206 ...
Page 207: ...207 Overview ...
Page 208: ...208 ...
Page 209: ...209 Overview ...
Page 210: ...210 ...