Sartoclear
®
Pilot Filter Holder
19
Operation
Moving the Pressure Plate and Clamping Plate to the Outermost Stop
Procedure
Depending on the number of filter cartridges to be inserted, the distances between
the pressure plate and clamping plate have to be adjusted in relation to each other .
NOTICE
Danger of damage to the hydraulic cylinder! Never pressurize the hydraulic
cylinder without inserted cartridges . Since the hydraulic cylinder only has a limited
stroke, the filter cartridges must be positioned very tightly together .
t
Using the holding clamps, position the plates on the pull rods:
t
Move the holding clamps on the clamping plate to the pull rod’s outermost
position (1) (see Chapter "5 .1 Moving Holding Clamps", page 17) .
t
Repeat the process with the second pull rod (2) .
t
Push the clamping plate (3) on to the holding clamps .
Moving the Pull Rods to the Outermost Stop
Procedure
t
Move the outer holding clamps (1) on the two pull rods to the respective
left-hand position (see Chapter 5 .1, page 17) .
t
Move both pull rods (2) as far as the stop on the right at the same time .
t
Move the inner holding clamps on the two pull rods to the respective left-hand
position (3) .
Inserting the Cartridges
Procedure
t
NOTICE
Danger of damage to the filtration system! If the manifold plate is
aligned incorrectly, the flow rate is disturbed by the filtration process, which
may cause the filtrate to leak or cause damage to the manifold plate and filter
cartridges .
t
Place the manifold plate (1) for the filtered solution inlet according to its
labeling (see operating instructions for the manifold plate) onto the bottom
guide rods .
t
The channels (2) must be aligned in the direction of the filter cartridges .
t
Push the manifold plate onto the pressure plate .
1
2
3
2
1
3
1
2
2