Sartoclear
®
Pilot Filter Holder
23
Malfunctions
6 Malfunctions
Problem
Cause
Correction
Chapter, page
The filter holder is not
leaktight .
The tensioning pressure is
insufficient
.
Check the pressure gauge display .
If required: Increase the inlet pressure .
Observe the tensioning force (kN) of the inserted
filter cartridge .
Operating instructions
for the filter cartridge
The hydraulic piston has
moved out completely
without the filter
cartridges being
sufficiently compressed .
Set the fine adjustment to a narrower distance .
Check whether the maximum permissible stroke
is exceeded .
Hydraulic oil is leaking
from the hydraulic
cylinder .
The hydraulic cylinder is
faulty .
Contact Sartorius Service .
Consult your Sartorius Service regarding any problems you cannot solve yourself .