8. Calibration of the MD8 airscan
As already described above, you can call up the calibration menu with a two-key combination
when the MD8 airscan is in the initial starting position, i.e. the message “sartorius MD8 ready
to start” is displayed.
Press the
⇑
and
⇓
key simultaneously for 3 seconds.
The calibration function is shown in the display by the message
“calibration – 3 point (START)”.
When
Start
is now pressed, the 3-point calibration mode is started.
To perform calibration, a Sartorius Stedim Biotech calibration unit 16740 must be connected.
Exact details on how to do this are given in section 5 of the operating instructions provided
with the calibration unit (publication number SL-6051a).
A complete reference curve for the flow rate sensor is stored in the MD8 airscan software,
which can be brought into line with each instrument by the 3-point calibration. In the
3-point mode, the air flow rates at the 3, 5 and 7 m
3
/h settings are successively run and
compared with the values measured by the calibration unit.
Calibration can be interrupted at any time by pressing the
Stop
key, and is done so
automatically when the unit becomes overheated. The points measured and recorded up
to this interruption are not stored, however. The last calibration values remained valid.
| 29
Summary of Contents for MD8 airscan
Page 1: ...85030 511 25 Operating Instructions Air Sampler MD8 airscan Ver 05 2011...
Page 2: ...Read these operating instructions carefully before switching the instrument on 2...
Page 18: ...Fix a gelatin filter unit to the adapter Fig 8 In principle sampling can now be started 18...
Page 36: ...Fig 14 Fig 15 36...
Page 37: ...Fig 16 Fig 17 37...
Page 38: ...Fig 18 38...
Page 40: ...Fig 19 40...
Page 52: ...17655 17656 17657 17657 17657 17208 52...
Page 53: ...17085 17088 17662 16970 16970 16970 53...