7.4. Display functions during air sampling
Throughout air sampling, the current air flow rate and the remaining time are continuously
updated on the display (“airflow #.# m
3
/h timer ## min”). Moreover, you can also press the
⇒
key to check the nominal settings during air sampling.You simply use this key to recall
each parameter the same way as when checking or changing the individual parameters,
one after the other. The only difference is that you cannot change the parameter settings
during air sampling.
This display function does not affect the air sampling procedure which is in progress. The
instrument automatically returns to the mode for displaying the current air flow rate and the
remaining time 10 seconds after a key was last pressed.