Sartorius Stedim Biotech 16756 Operating Instructions Manual Download Page 7

7

(See also Operating Instructions for
MD8 airscan Air Sampler)

A complete reference curve for 
the flow rate sensor is stored in the
MD8 airscan software, which can be
brought into line with each instru-
ment by the 3-point calibration. 
In the 3-point mode, the air flow
rates at the 3, 5 and 7 m

3

/h settings

are successively run and compared
with the values measured by the
calibration unit 16756. 

Calibration can be interrupted at
any time by pressing Stop. When
the unit becomes overheated, cali-
bration is interrupted automatically. 

The points measured and recorded
up to this interruption will be
remain stored and the unchanged
points are also stored.

You can call up the calibration
menu when the MD8 airscan is in
the initial starting position, i.e. 
the message “Sartorius MD8 ready
to start” is displayed, by simultane-
ously pressing the 

and 

keys

for 3 seconds.

The calibration function is shown in
the display by the message: 
“– Calibration – 3 point (Start)”.

When Start is pressed with the unit
in this position, the 3-point calibra-
tion mode is started.

6.1 Calibration Procedure

(After connection of the calibration
unit is completed as described in
Section 5). 

a) The MD8 airscan moves to the first

measurement point, and continual-
ly displays the nominal and actual
values
(The message in the display reads:
“Nominal 3.0 m

3

/h Actual: 

xx m

3

/h”).

b) As soon as the nominal value is

reached and stabilized, which can
be recognized from the actual value
display, press the parameter key 

.

c) The MD8 airscan now continues to

run uncontrolled, i.e. the set point
of the motor does not change, 
so that the air flow rate remains
constant. In this mode, the display
reads: 
“Nominal 3.0 m

3

/h real: x.xx m

3

/h”. 

d) As soon as the connected calibra-

tion unit 16756 has stabilized and
displays a constant value, set this
value on the MD8 airscan by press-
ing 

or 

. Then confirm and

store it with Enter.

e) If this is not the last calibration

point, the next one is automatically
activated and processed as
described under a–d. 
After the 7 m

3

/h value has been cal-

ibrated, the calibration is finished. 
This is shown by the message 
“Calibration completed (Enter)”.

f) Confirm by pressing Enter. 

The MD8 airscan returns to the 
initial starting position and is 
ready to start.

Note:
If the calibration is interrupted
ahead of time, either the message
“User break! 
(press Enter)” or “Overheated device
(Enter)” will be displayed. Confirm
this message by pressing Enter. 
The MD8 airscan returns to the 
initial starting position.

The altered calibration curve 
is stored immediately when the 
calibration has been finished or
interrupted.

6

Calibration of the MD8 airscan:

6.2. Resetting of Default Values

If, when calibrating the MD8
airscan, data were entered which
are so absurd that they make it
impossible for the instrument to be
controlled correctly by the phase
controller, it may be very difficult
to recalibrate correctly.

Even though this case is very
unlikely, because the consequences
would be serious, a hidden option
has been implemented in the MD8
airscan, allowing it to be reset to 
its default values.

Procedure:
Switch off the MD8 airscan. 
While holding down both the 

and Stop keys simultaneously and
switch the MD8 airscan on. Do not
release the keys until the message
“Self-testing” is displayed. 
Now press the 

key while this

message is displayed.

Pressing of the 

key must be

made while the “Self-testing” 
message is displayed as only then
will the display change to “Delete
data? <Start> = yes”. 
By now pressing Enter you confirm
this and all data will be overwritten
to the default settings. A self-test is
then re-started. 

If you press any other key, instead
of Enter, the MD8 airscan will enter
its standard operating mode with-
out altering the data, in other
words, the unit is reset to the
default values.

Summary of Contents for 16756

Page 1: ...Operating Instructions Bedienungsanleitung Calibration Unit 16756 Kalibriereinheit 16756 85030 523 26 Vers 11 2011...

Page 2: ...alibration Unit 3 3 Specifications 4 4 Equipment Supplied 5 5 Operation of the Calibration Unit 5 6 Calibration of the MD8 airscan 7 7 Calibration of the AirPort MD8 8 8 Care and Maintenance 9 9 Infor...

Page 3: ...ont panel Fig 1 see a and an air outlet port on the rear panel see b for attaching an 17085 hose for connection to the air sampler bayonet design quick connect Located on the right side Fig 1 c is the...

Page 4: ...he battery charger without overloading the rechargeable battery pack The unit can also be used when recharging run on mains power supply We recommend leaving the calibration unit connected to the batt...

Page 5: ...one on the MD8 airscan AirPort MD8 air sam plers See the operating instructions for the MD8 airscan AirPort MD8 air samplers regarding preparation of the filter holder placing the filter in the filte...

Page 6: ...uring calibration no filter holder is attached to the air sampler Attach the hose to the end of the connector Fig 4 and attach the 17659 connector hose calibration unit to the free end of the hose Now...

Page 7: ...alue set this value on the MD8 airscan by press ing or Then confirm and store it with Enter e If this is not the last calibration point the next one is automatically activated and processed as describ...

Page 8: ...flow rate 7 3 Calibration Procedure After connecting the calibration unit and pressing the Start button on the calibration unit the message Sartorius MD8 Calibrating System start air sampler appears i...

Page 9: ...in your regular household waste such equipment is manu factured from high grade materials and can be recycled and reused The European Directive 2002 96 EC on Waste Electrical and Electronic Equipment...

Page 10: ...ax 44 1372 726171 In countries that are not members of the European Economic Area EEA or where no Sartorius affili ates subsidiaries dealers or distrib utors are located please contact your local auth...

Page 11: ...not differentiate between still utilizable and or repairable used electrical and electronic equipment and no longer utilizable and or repairable used electrical and electronic equipment or between was...

Page 12: ...er tebeschreibung 13 3 Technische Daten 14 4 Lieferumfang 15 5 Inbetriebnahme des Ger tes 15 6 Kalibrierung des MD8 airscan 17 7 Kalibrierung des AirPort MD8 18 8 Pflege und Wartung 19 9 Entsorgungshi...

Page 13: ...er das Ger t unter Umgehung der Abschaltautomatik ausschaltet die Kalibriereinheit besitzt keinen Ausschalter die Kalibriereinheit schaltet sich nach 1 Minute selbst t tig ab Auf der Oberseite der Kal...

Page 14: ...Max Fehler des Messsystems 2 1 m3 h 16 m3 Messsignal 2500 Impulse 1 m3 Messbereich der Frequenzmessung 0 6 Hz bis 12 Hz Genauigkeit 0 005 Hz Automatische Abschaltung nach 1 Minute bei Frequenz 0 4 Hz...

Page 15: ...Filters und Verschlie en des Filterhalters erfolgen entsprechend der Bedienungsanleitungen f r die MD8 airscan AirPort MD8 Lufkeimsammelger te in denen auch die Handhabung Gebrauch der Gelatinefilter...

Page 16: ...60 Sekunden sollte jetzt mit der Kalibrierung begonnen werden Geschieht dies nicht schal tet sich das Ger t nach 60 Sekunden ab Durch erneutes Dr cken der gr nen Starttaste kann das Ger t wieder einge...

Page 17: ...end best tigen und spei chern Sie den Wert durch Bet tigen der Enter Taste e Wenn dies noch nicht der letzte Kalibrierwert ist m ssen Sie jetzt den n chsten Kalibrierwert wie unter a d beschrieben anf...

Page 18: ...librierung Nach Anschluss der Kalibriereinheit und Dr cken der Starttaste an der Kalibriereinheit erscheint Sartorius MD8 Calibrating System start air sampler Am AirPort MD8 wird nun aus dem Men punkt...

Page 19: ...m ll denn sie sind aus hochwertigen Materialien hergestellt die recyclet und wieder verwendet werden k nnen Die europ ische Richtlinie 2002 96 EG WEEE fordert die elektrischen und elektronischen Ger t...

Page 20: ...34 91 3589623 Ungarn Sartorius Stedim Hung ria Kft Kagyl u 5 2092 Budakeszi Telefon 36 23 457 227 Fax 36 23 457 147 In den L ndern die keine Mitglieder des Europ ischen Wirtschaftsrau mes sind oder i...

Page 21: ...u betr chtlichen Verz gerungen in der Abwicklung Bei der R cknahme wird nicht zwischen noch verwendungs bzw reparaturf higen Elektroaltger ten und nicht mehr verwendungs bzw reparaturf higen Elektroal...

Page 22: ...ermany All rights reserved No part of this publication may be reprinted or translated in any form or by any means without the prior written permission of Sartorius Stedim Biotech GmbH The status of th...

Reviews: