Appendix 45
9.10 Error reports and Warning messages
The following tables list all error reports and warning messages which can appear in
the operator control .
Error report
Possible cause
Remedy
All runs blocked (X)
(X) = Error number
No. Error
0
Turning axis
1
Turning squeeze axis
2
Fixed squeeze axis
3
Cutting axis
4
Forward axis
5
Heating*
6
Fan
7
Placeholder
8
Placeholder
9
Cover lock
Run stopped by power fail
see Chapter 4 .5 Blocked device
Follow the instructions on the user interface .
* The (5) Heating Error is triggered when the blade temperature has not reached the depyrogenisation temperature of < 400°C
Warning messages
Possible cause
Elimination
Please remove tube holder
Tube holder inserted
Tube holder removed
No tube holder inserted
1 . No tube holder inserted
2 . Contact surface of tube holder or the
spring contacts are dirty .
1 . Insert tube holder set
2 . Clean the contact surfaces of the tube
holders and those of the spring contacts with
a suitable cleansing agent
No matching tube was found
No welding parameters available for the
selected type of tube and the inserted tube
holder set .
Was the right type of tube chosen?
Request new welding parameters from Sartorius
Stedim Biotech .
Only one tube holder inserted
1 . Only one tube holder is inserted
2 . Only one tube holder is recognized .
1 . Insert both tube holders .
2 . Clean the contact surfaces of the tube
holders and those of the spring contacts with
a suitable cleansing agent
Tube holders do not match
Tube holders with different color marking
were inserted .
Insert tube holders with the same color marking .
Remove blade
Blade has not yet been removed up to now
Remove blade
Blade not inserted
Blade not completely inserted to back
of holder
Insert blade