― 38 ―
― 37 ―
5-8
「
」
(Max. rated input voltage: 50 mV dc)
・
Temperature measurement (
℃
) or (
o
F )
( For K-type thermocouple)
1. Pay attention in order to avoid risk of burn depending on the
object temperature or measuring environment.
2. Do not apply exceeding 50 mV to the measuring terminals.
WARNING
1) What to measure
℃
,
o
F (Temperature): Temperature of liquid, solids, gas, and
etc.
2) Measuring ranges
Celsius: -50.0
℃ ~
1000.0
℃
Fahrenheit: -58.0
o
F
~
to 1832.0
o
F
3) Measuring procedure
①
Connect the provided K-type thermocouple to the Temp
measuring terminals.
②
Set the function selector to Temp.
③
Press the SELECT button to select
℃
or
o
F.
④
Apply the thermocouple to the object to measure.
⑤
Read the display.
Note:
•
Temperature function does not show the bar graph.
•
The provided K-type thermocouple (K-250PC) is a polar device.
Connect the device to the meter properly.
•
The range of K-250PC is -50
℃
to 250
℃
.
•
Separately available K-type adapter (K-AD) allows you to use
other international standard mini plug thermocouples.
Temp
Temp