― 66 ―
― 65 ―
DC Voltage/AC Voltage, DC+AC Voltage/AC Voltage for dual display
Range
Accuracy*
DC, 45 Hz
~
65 Hz
500.00 mV
5.0000 V
50.000 V
500.00 V
1000.0 V
Main display:± (0.7 % rdg + 60 dgt)
Sub display:± (0.7 % rdg + 6 dgt)
65 Hz
~
500 Hz
500.00 mV
Main display:± (1.0 % rdg + 40 dgt)
Sub display:± (1.0 % rdg + 4 dgt)
5.0000 V
50.000 V
500.00 V
Main display:± (1.2 % rdg + 60 dgt)
Sub display:± (1.2 % rdg + 6 dgt)
1000.0 V
Main display:± (1.7 % rdg + 60 dgt)
Sub display:± (1.7 % rdg + 6 dgt)
500 Hz
~
1 kHz
500.00 mV
Main display:± (1.0 % rdg + 40 dgt)
Sub display:± (1.0 % rdg + 4 dgt)
5.0000 V
50.000 V
500.00 V
1000.0 V
Main display:± (1.7 % rdg + 60 dgt)
Sub display:± (1.7 % rdg + 6 dgt)
1 kHz
~
20 kHz
500.00 mV
±1 dB**
5.0000 V
50.000 V
±2 dB**
500.00 V
±3 dB**
1000.0 V
Unspecified
Input impedance: 10 MΩ, 60 pF nominal
(80 pF nominal for 500.00 mV range)
Residual reading: Less than 50 digits with test leads shorted.
* From 5 % to 10 % of the range: ±(Specified % of rdg + 80 dgt)
** From 5 % to 10 % of the range: ±(Specified % of rdg + 180 dgt)
From 10 % to 15 % of the range: ±(Specified % of rdg + 100 dgt)
AC Voltage (with low-pass filter for Variable Frequency Drive)
Range
Accuracy*
10 Hz
~
40 Hz
5.0000 V
50.000 V
500.00 V
1000.0 V
± (3.5 % rdg + 80 dgt)
40 Hz
~
200 Hz
5.0000 V
50.000 V
500.00 V
1000.0 V
± (2.0 % rdg + 60 dgt)
200 Hz
~
440 Hz
5.0000 V
50.000 V
500.00 V
1000.0 V
± (7.0 % rdg + 80 dgt)**
*Not specified for fundamental frequency > 440 Hz
**Accuracy linearly decreases
from ±(2.5 % of rdg + 60 dgt) @ 200 Hz
to ±(7.0 % of rdg + 80 dgt) @ 440 Hz.
Decibel dBm
Range and accuracy are subjected to ACmV, ACV, and reference
impedance selected.
Typical 600 reference impedance ranges:
ACmV: -29.83 dBm
~
-3.80 dBm
ACV: -9.82 dBm
~
54.25 dBm
Accuracy:±0.25 dB + 2 dgt (40 Hz
~
20 kHz)
Input impedance: 10 MΩ, 60 pF nominal
Selectable reference impedance:
4, 8, 16, 32, 50, 75, 93, 110, 125, 135, 150, 200, 250, 300, 500, 600,
800, 900, 1000, 1200 Ω