Frequency (Hz)
Range
Input sensitivity*
Frequency range***
500.00 mV
100 mV
10.00 Hz
~
200.0 kHz
5.0000 V
0.6 V
10.00 Hz
~
100.0 kHz
50.000 V
6 V
500.00 V
50 V
1000.0 V
500 V
10.00 Hz
~
10.00 kHz
LPF 5.0000 V
0.5 V
~
2 V**
10.00 Hz
~
440.0 Hz
LPF 50.000 V
5 V
~
20 V**
LPF 500.00 V
50 V
~
200 V**
LPF 1000.0 V
500 V
~
1000 V
10.00 Hz
~
200.0 Hz
500.00 μA
50 μA
15.00 Hz
~
10.00 kHz
5000.0 μA
500 μA
10.00 Hz
~
10.00 kHz
50.000 mA
5 mA
500.00 mA
50 mA
5.0000 A
1 A
10.00 Hz
~
3.000 kHz
10.000 A
8 A
Accuracy: Sub display ±0.02 % rdg+ 4 dgt
Main display ±0.02 % rdg+ 40 dgt
* Specified based on sine wave RMS
** LPF sensitivity linearly decreases
from 10 % of F.S. @ 200 Hz
to 40 % of F.S. @ 440 Hz.
*** Frequency ranges on the Sub display
A Least Significant Digit will be additionally displayed on the
Main display.
Logic level frequency (
) and Duty cycle (D%)
DCmV function
Range
Accuracy*
Frequency
5.000 Hz
~
2.0000 MHz ±
(
0.002 % rdg + 4 dgt
)
Duty cycle
0.1 %
~
99.99 % ±
(
3 dgt/kHz + 2 dgt
)
**
* Sensitivity: 2.5 Vp (Square wave) for 3 V and 5 V logic family
** Frequency range: 5 Hz to 500 kHz
Max/Min (with capture mode)
Resolution: Equivalent to 5000 counts
Accuracy: ±(Specified % of rdg + 100 dgt) in each function
Sampling time: Approx. 0.8 ms
Max/Min (with recording mode)
Accuracy: Specified accuracy in each function
How to calculate an accuracy
Example) DC voltage measurement (DC mV)
Reading: 100.00 [mV]
Range accuracy: ±(0.03 % rdg+2 dgt) in the 500.00 mV range
Measuring error: ±(100.00 [mV]×0.03 % rdg + 2 dgt)
≒
±0.05 [mV]
Calculation: 100.00 [mV] ±0.05 [mV]
True value
: from 99.95 [mV] to 100.05 [mV]
* 2 dgt in the 500.00 mV range corresponds to 0.02 mV.
The product specifications and its appearance described in
this manual are subject to change without prior notice for
improvements or other reasons.
Capacitance
Range
Accuracy*
50.00 nF
500.0 nF
±
(
0.8 % rdg + 3 dgt
)
5.000 μF
±
(
1.5 % rdg + 3 dgt
)
50.00 μF
±
(
2.5 % rdg + 3 dgt
)
500.0 μF**
±
(
3.5 % rdg + 5 dgt
)
5.000 mF**
±
(
5.0 % rdg + 5 dgt
)
25.00 mF**
±
(
6.5 % rdg + 5 dgt
)
* Accuracies with film capacitor or better
** In manual-ranging mode, measurements not specified below
45.0 μF, 0.450 mF and 4.5 mF for 500.0 μF, 5.000 mF and
25.00 mF ranges respectively