― 34 ―
― 33 ―
•
The display style of [m
/Hz] shows the bar graph.
•
As a normal condition, non-connected test leads may cause
unstable readings.
•
Manual trigger level selection on Hz reading is not available in
this function.
Note:
Range
Frequency measurement (Hz)
Input sensitivity (Sine wave)
Frequency range
500.00 mV ac
100 mV
10.00 Hz
~
200.0 kHz
5-7
「
」
(Do not apply any voltage or current.)
・
Resistance (Ω) measurement
・
Conductance (nS) measurement
・
Continuity check ( )
1) What to measure
・
Ω
(Resistance): Resistor, circuit resistance, etc.
・
(Continuity check): Wiring connections, Operation of
switches, etc.
・
nS(Conductance): High-value resistance of Giga-Ohms for
leakage measurements
Note: Conductance is the inverse of Resistance, that is S=1/Ω
or nS=1/GΩ.
2) Measuring ranges
・
Ω
: 6 ranges; 500.00 Ω, 5.0000 kΩ, 50.000 kΩ, 500.00 kΩ,
5.0000 MΩ, and 50.000 MΩ
・
: Beeper threshold level: between 20 Ω and 200 Ω,
Response time: <100 μs
・
nS : 99.99 nS (Single range)
*Open circuit voltage between the measuring terminals: <1.3 V dc
(<3 V dc for 500.00 Ω range)
Do not apply any voltage or current to the measuring terminals.
In the case of high resistance measurement, readings may be
unstable due to external inductive influence.