― 68 ―
― 67 ―
DC current
Range
Accuracy
Input resistance**
500.00 μA
±(0.15 % rdg + 20 dgt)
Approx. 100 Ω
5000.0 μA
±(0.1 % rdg + 20 dgt)
50.000 mA
±(0.15 % rdg + 20 dgt)
Approx. 1 Ω
500.00 mA
±(0.15 % rdg + 30 dgt)
5.0000 A
±(0.8 % rdg + 20 dgt)
Approx. 0.01 Ω
10.000 A*
±(0.5 % rdg + 20 dgt)
* > 6 A: Cool down more than 3 minutes after measuring 1 minute.
< 6 A Continuable
**Fusing resistor not included
DC loop current % 4
~
20 mA
4 mA=0 % (zero), 20 mA=100 %(span)
Resolution: 0.01 %, Accuracy:±25 dgt
DC current/AC current and DC+AC current/AC current
Range
Accuracy
Input
resistance**
DC, 50 Hz
~
60 Hz
500.00 μA
Main display:±(0.6 % rdg + 40 dgt)
Sub display:±(0.6 % rdg + 4 dgt)
Approx. 100 Ω
5000.0 μA
50.000 mA
Approx. 1 Ω
500.00 mA
5.0000 A Main display:±(1.0 % rdg + 40 dgt)
Sub display:±(1.0 % rdg + 4 dgt)
Approx. 0.01 Ω
10.000 A*
40 Hz
~
1 kHz (Except 50 Hz
~
60 Hz)
500.00 μA
Main display:±(1.0 % rdg + 50 dgt)
Sub display:±(1.0 % rdg + 5 dgt)
Approx. 100 Ω
5000.0 μA
50.000 mA
Approx. 1 Ω
500.00 mA
5.0000 A
Approx. 0.01 Ω
10.000 A*
* > 6 A: Cool down more than 3 minutes after measuring 1 minute.
< 6 A Continuable
**Fusing resistor not included
Resistance Ω
Range
Accuracy
500.00 Ω
±(0.2 % rdg + 10 dgt)
5.0000 kΩ
±(0.2 % rdg + 6 dgt)
50.000 kΩ
500.00 kΩ
5.0000 MΩ
±(0.8 % rdg + 6 dgt)
50.000 MΩ
±(2.5 % rdg + 6 dgt)
99.99 nS*
±(1.0 % rdg + 10 dgt)
Open circuit voltage: <1.3 V dc (<3 V dc for 500.00 Ω range)
*From 0 % to 10 % of the range: ±(Specified % of rdg + 30 dgt)
Continuity check
Threshold level: 20 Ω to 200 Ω
Response time: < 100 μs
Diode test
Range
Accuracy
Test current
Open circuit
voltage
2.0000 V ±
(
1.0 % rdg +10 dgt
)
Approx. 0.4 mA
< 3.0 V
Temp (
℃
&
o
F)**
Range
Accuracy*
-50.0
℃~
-10.0
℃
±(0.3 % rdg + 30 dgt)
-10.0
℃~
1000.0
℃
±(0.3 % rdg + 20 dgt)
-58.0
o
F
~
14.0
o
F
±(0.3 % rdg + 60 dgt)
-14.0
o
F
~
1832.0
o
F
±(0.3 % rdg + 40 dgt)
* Accuracy with K-type thermocouple.
K-type thermocouple range and accuracy not included
**Cool down more than 30 minutes after measuring DCA or ACA.