6-7-1 C1,C2 ranges
1) Application
Measurement of capacitance
2) Measuring range
C1 range : 50 pF
〜
0.2 µF
C2 range : 0.01
〜
20 µF
3) Measurement procedure
①
Shift the polarity reversal switch to
+
position.
②
Connect the black plug of the test lead to the
−
input
terminal and the red plug to the
+
input terminal.
③
Set the function/range selector to an appropriate C1 or
C2
range.
④
Push the function power switch. (See 5-1
④
)
Then, the power indicator blinks.
⑤
Short the test pins and turn the 0
Ω
・
C
∞
adjuster to have
the pointer align exactly with
∞
of C1or C2 scale.
⑥
Apply the black and red test pin to the measured capacitor.
6-7 Capacitance Measurement ( )
1.Never apply voltage to the input terminals.
2.Discharge the capacitance before measuring it.
WARNING
⑦
Read the pointer on C1or C2 scale.
⑧
After measurement, remove the red and black test pins from
the object measured.
⑨
Turn off the function power switch. (See 5-1
④
)
⑨
⑤
②
①
③
⑧
⑧
Capacitor
②
⑥
④
⑦