You can also use the SantaMedical RY210  to measure the temperature
of an area or an object, a food, a liquid or a room temperature.
For this, use the SURFACE TEMP mode.
Measurement range for Surface TEMP mode: 0

ć

-60

ć

 (32

̧

-140

̧

)

Press PROG SETTING button for 3 seconds, the screen displays: F1.
Press PROG SETTING button twice to get F3.
Select 0 by pressing “-” button to get the BODY mode.
Select 1 by pressing “+” button to get the SURFACE TEMP mode
Confirm by pressing PROG SETTING button.

5.TATAL DIFFERENCE – F4 MANU

          

 

Press PROG SETTING button for 3 seconds, the screen displays: F1.
Press PROG SETTING button three times to get F4.
Select “+” to increase the difference by 0.1

ć

 (0.1

̧

), “-” to reduce it by 

0.1

ć

 (0.1

̧

).

Confirm by pressing PROG SETTING button.
If  in  doubt,  you  are  advised  to  leave  the  total  variance  at  +  0.0

ć

 

(original setting).

Each time there is a significant change in the ambient temperature due 
to a change in environment, to allow the RY210 to acclimatize to 
this ambient temperature for at least 15 minutes before using it. 

Note: The SantaMedical RY210  is automatically set to BODY

Important:  The  area  temperature  differs  from  the  internal  body 
temperature.  To  obtain  the  internal  temperature  always  use  the 
BODY mode.

Please  make  sure  to  select  the  BODY  mode  for  an  internal 
temperature reading and the SURFACE TEMP mode for an external 
area reading (bottle, bath, Room…).

 To adjust total variation of your SantaMedical RY210 

Summary of Contents for RY210

Page 1: ...Model RY210 Non contact Infrared Thermometer ...

Page 2: ...CONSIRAINTS DESCRIPTION OF THE RY210 FUNCTIONS USAGE SETTING FUNCTION OF MENU 1 SWITCH ON THE DEVICE 2 CHOOSING THE TEMPERATURE UNIT F1 MENU 3 ALARM SETUP F2 MENU 4 CHOICE OF DISPLAY MODE BODY OR SURFACE TEMP F3 MENU 5 TOTAL DIFFERENCE F4 MENU 6 BUZZER ON OFF F5 MENU 7 EXITING THE SETTING MODE 8 DATA MEMORY 9 CHANGING THE BATTERIES TECHNICAL CHARACTERISTICS PRECISION ADVICE ACCESSORIES SUPPLIED TR...

Page 3: ......

Page 4: ...the SantaMedical RY210 properly in order to obtain reliable and stable results You are therefore advised to read this instruction manual and the safety precautions carefully before use 1 Greenes D Fleisher G Accuracy of a Noninvasive Temporal Artery Thermometer for Use in Infants Arch Pediatr Adolesc Med 2001 155 376 III PRECAUTIONS BEFORE USE In order to obtain reliable and stable results you are...

Page 5: ...nd involves measuring the temperature in the pulmonary artery by means of a catheter equipped with a thermal probe which can read the temperature in situ The same method is employed for probes measuring the oesophageal temperature However such invasive temperature measurement methods require specific equipment and expertise Rectal thermometry Rectal temperature adjusts slowly in comparison to the ...

Page 6: ......

Page 7: ...is artery is linked to the heart by the carotid artery which is directly linked to the aorta It forms part of the main trunk of the arterial system The efficiency speed and comfort of taking a temperature from this area make it ideal compared with other temperature measurements methods NORMAL TEMPERATURES ACCORDING TO AGE PRACTICAL CONSIDERATIONS WHEN TAKING A TEMPERATURE In order to ensure that p...

Page 8: ...hen standing HOW TO TAKE A TEMPERATURE Aim at the FOREHEAD over the right temporal region from a distance of about 3 5cm 1 2in 2in press the thermometer s measurement button and the temperature is instantly Displayed The reliability of the measurement cannot be guaranteed if the temperature is measured over another part of the body e g arm torso CONSTRAINTS Please observe the following before any ...

Page 9: ...8 V DESCRIPTION OF THE SantaMedical RY210 Infrared Lens LCD screen Measuring key Battery compartment ...

Page 10: ...RY210 SantaMedical Non Contact Infrared Thermometer Model RY210 Operating range 3cm 5cm Power DC 3V RQVXPSWLRQ P SantaMedical 9 SETTING KEYS MENU BUTTON WEEE IEC60417 5840 Type B applied part TRADEMARK ...

Page 11: ...ath or room temperature by using the Surface Temp function This function is in accordance with the Directive 89 336 EEC Electromagnetic Compatibility VII USAGE 1 Install battery 2 For the first usage or when inserting new batteries wait between 10 minutes for the warm up of the apparatus and when inserting the new batteries 3 Aim towards the forehead see the diagram below for the RY210 positioning...

Page 12: ......

Page 13: ...mes to get F4 Select to increase the difference by 0 1ć 0 1 to reduce it by 0 1ć 0 1 Confirm by pressing PROG SETTING button If in doubt you are advised to leave the total variance at 0 0ć original setting Each time there is a significant change in the ambient temperature due to a change in environment to allow the RY210 to acclimatize to this ambient temperature for at least 15 minutes before usi...

Page 14: ... button You will obtain the last temperature measurement To switch on the last measurement press the button The number indicated in interval of two measures corresponds to the number of measurements To exit data memory press the measure key The SantaMedical RY210 switches off in 30 seconds 9 CHANGING THE BATTERIES Display when the LCD screen displays Battery the battery is used Operation Open the ...

Page 15: ...nce 3cm 5cm 1 2in 2in 10 Automatic stop 30 sec LONGEVITY USE The SantaMedical RY210 is guaranteed for 40 000 takings X ADVICE The protective glass over the lens is the most important and fragile part of the thermometer please take great care of it Clean the glass with cotton fabric wet with 95 alcohol Do not use other batteries than mentioned batteries do not recharge non rechargeable batteries do...

Page 16: ...ing the Program Setting button to get the F1 function key The screen displays the body temperature inferior to 32 89 6 To take a body temperature the function key F3 must be on Body mode If you re on Surface Temp mode the 32 89 6 temperature displayed is showing the external temperature that your body releases The screen displays the message HI When using the SantaMedical RY210 the message HI can ...

Page 17: ... LO message display Advice Make sure there is no air flux as this could interfere with the infrared system Make sure that there is no obstruction prior to taking a temperature Temperature hampered by an air flux Temperature readings too close together the RY210 did not have the chance to boot itself Respect the pause of 15 seconds minimum between two readings 3 5 seconds pause is advised Please re...

Reviews: