15
XII.TROUBLESHOOTING
If you have one of the following problems while using your SantaMedical
RY210, please refer to this breakdown service guide to help resolve
the problem. If the problem persists please contact our customer
service.
The screen displays temperature superior to 95°
The temperature is in Fahrenheit change the measurement to Celsius
(by pressing the Program Setting button to get the F1 function key)
The screen displays the body temperature inferior to 32
℃
(89.6
℉
)
To take a body temperature the function key F3 must be on Body
mode. If you’re on Surface Temp mode the 32
℃
(89.6
℉
) temperature
displayed is showing the external temperature that your body releases.
The screen displays the message HI
When using the SantaMedical RY210 the
message HI can show on the screen.
The analysis is above the measurement range
selected, either superior to 42.9
℃
(109.2
℉
)
in Body Mode or superior to 60
℃
(140
℉
) in
Surface Temp Mode.