Sanosil Q-Jet Advanced
17
•
Shortening of waiting time: after a
minimal waiting time of 120 minutes
place a strong dehumidifier in the
middle of the room and dehumidify room for at least 60 minutes or air room by opening a window or HVAC. Use
full protection equipment when placing dehumidifier!
•
When room re-entry is required before treatment process is completed or to perform a H
2
O
2
level check, proper
personal equipment (PPE) including wraparound style goggles; respirator mask (filter class ABE1) filtering hydrogen
peroxide (for example: 3M 4277 or 3M 7500 mask with 6057 filter) and gloves should be worn to safely enter the room.
WARNING!
Risk of injuries!
To prevent injuries do always check H2O2-Level before entering the room.
When entering room to measure
the H2O2-Value, do always wear full protection equipment, i.e. protection mask, safety googles and gloves.
The disinfected room can only be entered without protection equipment if the H2O2-Value is below 1ppm.
•
Tip ON/OFF (3) to switch off device