Sanosil Q-Jet Advanced
16
TIME: 30:00
5.4
Automatic (with timer)
Please read and follow important safety directions on page 6
before use!
•
Choose AUTOMATIC mode and confirm by pressing ENTER
•
Pre-set nebulization time is displayed
•
For pre-setting device (set time prior use),
please see below,
SETUP TIMER
•
Confirm pre-settings by pressing ENTER
(active = black dot on display)
CAUTION
If AUTOMATIC-mode is
„ON“
and disinfection process is set by the weekly-timer, the nebulization time
cannot be over 1 (one) hour! Please make sure that during fogging and after the required waiting
time no living being are allowed to be in the treated room.
SETUP TIMER:
•
Choose mode SETUP AUTO and confirm by pressing ENTER
•
Enter desired values
-
MC = Volume of room in m3
-
CONC = Concentration (dosage) in ml per m3
Nebulization TIME will be calculated automatically
For the approx. fogging time and consumption refer to the table on page 14
•
Weekly timer (press arrow button „down“) (activated days
are shown through filled squares)
•
To program days press ENTER
- For each day input starting-time and confirm with ENTER
•
To
activate day press “I”
-button
(square is filled when day is activated)
•
Device will start at the times programmed
Please make sure there are no living beings inside the room!
•
During the course of the disinfection program the display shows the remaining time (Count-down)
•
At the end of the disinfection program the green LED flashes.
•
Room re-entry minimum waiting-time begins (minimal 120 Minutes)
Recommended minimal “
Nebulization time / Minimum waiting time
”:
0
–
20min/120 min, 20
–
30 min/130min, 30
–
40 min/140min, 40
–
50/min/150min, 50
–
60min and above 160min
(a overnight treatment is recommended, see 2.7)
•
After the end
of the waiting-time the room can be prepared for re-entry :
•
Check with a hydrogen peroxide level sensor that the H2O2-Level is below 1ppm
•
Open windows and doors for min. 15 minutes, unseal HVAC vents and restart HVAC system to allow increased airflow
circulation in the room. Once a safe ppm level is reached, room can be re-used again.
Cont. next page
TIME: 30:00
AUTOMATIC
START
WED
HH:xx:xx
MC:180
START
WED
HH:02:30
MC:180
SETUP AUTO
MC: xxx CONC: xx
TIME: xx:xx
M T W T F S S