34
6.3
Technical specifications
Technical specifications
Semiconductor version
Electrochemical version
Infrared version
Power supply voltage **
24VDC/AC +/- 20%, 5W , 50/60Hz
User interface
App with Bluetooth
Analogue output:
4-20mA / 0-10V / 1-5V / 2-10V selected via software
Serial communication:
Modbus® RS485 isolated slaves
Digital output 1 SPDT:
Alarm - relay 1 A/24 VDC/AC, resistive load
Digital output 2 SPDT:
Warning/FAULT - relay 1 A/24 VDC/AC, resistive load
Relay failsafe
Yes, selectable
Selectable delay:
0-20 min; 1-minute steps, selectable via Modbus register/app
Hysteresis
± 10% of the threshold value
IP protection:
IP67
Typical operating range:
0-1000 ppm
0-4000 ppm
0-100 ppm
0-10000 ppm
Sensing element
Pre-calibrated (also available as a spare part) with certificate
Remote cable length
5 metres
Storage temperature
-40 °C to +50 °C.
Storage humidity
5-90% relative humidity, non-condensing.
Storage position
Any
Operating temperature
-40 °C to +50 °C.
Operating humidity
5-90% relative humidity, non-condensing.
Maximum installation altitude
2000 metres
Operating position
Intended for vertical mounting with the sensor at the bottom
Precision*
<-10%/+15%
±5%
±5%
Start-up time*
5 minutes
5 minutes
2 minutes
Working life *
5 years
2 years
7 years
Calibration procedure requirements
12 months
12 months
Not required
*Reference conditions at 25°C 50% RH atmospheric pressure 101.3 kPa
** The device is intended to be supplied from an isolated Limited Energy Source per UL61010-1, 3rd edition cl. 9.4 or
Limited Power Source per UL60950-1 or Class 2 per NEC
6.3.1
Mechanical specification
Dimensions
Enclosure size (W
×
H
×
D) (approx.)
Built-in: 233x175x97 mm
Remote: 233x175x97 mm
Product casing (approx.)
Built-in: 590 g
Remote: 850 g
Summary of Contents for GLACIAR MIDI
Page 1: ...GLACI R MIDI Gas Leakage Detector for commercial and industrial applications USER MANUAL...
Page 5: ...5 2 2 Physical dimensions Built in version Remote version...
Page 29: ...29 5 4 1 Diagram of the calibration procedure...
Page 30: ...30 5 4 2 Operation of the registers for calibration...
Page 38: ...38...
Page 39: ...39...
Page 40: ...40 Manufactured by Samon AB Modemgatan2 S 23539Vellinge Sweden www samon se...