Supply the gas at the known concentration using the
calibration kit.
Wait about 1 minute until the gas concentration stabilises.
Click Span Calibration to set the calibration
concentration.
5
6
Enter the room temperature and relative humidity.
These values will be entered on the calibration certificate to
indicate the environmental conditions during calibration.
It is not necessary to use a calibrated instrument to perform
this measurement, an indicative value is sufficient.
Verify the summary screen, checking that all the
information has been entered correctly before
generating the calibration report.