25
5
MAINTENANCE
5.1
Calibration procedure
The calibration procedure is performed periodically and involves introducing a known gas concentration at the sensor
inlet, using the calibration kit.
The need to perform calibration is signalled by a specific variable on the supervisor. Each type of device has a different
calibration interval, as described in the technical specifications table. After a few years of operation, the sensor needs to
be replaced, as described in the following chapters, as the calibration is no longer sufficient to guarantee reliability of
the measurement performed.
The CO2 detectors do not require periodic calibration, but simply replacement of the sensor after approximately 7 years.
Calibration can be performed every 12 months if wanting to guarantee greater accuracy of the measurement or if needing
to issue a new calibration certificate. Below is a description of how to calibrate via the controller or via the app.
5.2
Calibration kit
The calibration kit is used to perform periodic calibration required for maintenance of the device. The gas cylinder and
pressure adapter to perform the calibration need to be procured separately.
Calibration kit with adapter, humidifier
Use the calibration adapter supplied
Wet
the filter core with tap water
Put the core back in place and close the filter. Note the direction of air flow (towards the
sensor).
Open the regulator on the gas cylinder and let the gas flow for approximately one minute
without the calibration adapter connected to the sensor, then close the regulator.
Summary of Contents for GLACIAR MIDI
Page 1: ...GLACI R MIDI Gas Leakage Detector for commercial and industrial applications USER MANUAL...
Page 5: ...5 2 2 Physical dimensions Built in version Remote version...
Page 29: ...29 5 4 1 Diagram of the calibration procedure...
Page 30: ...30 5 4 2 Operation of the registers for calibration...
Page 38: ...38...
Page 39: ...39...
Page 40: ...40 Manufactured by Samon AB Modemgatan2 S 23539Vellinge Sweden www samon se...