54
REV. 03-SIE/SRE
Example 9A, 9B, and 9C:
Selecting Channels
If a user simply wants to pre-configure his meter for
three or four pipe configurations, then it is a simple
matter of selecting the channel he is interested in by
changing the value of item 701 (Chan Select) in the
System Setup Menu. If on the other hand, the user
wants a generic meter, calibrated for Velocity Units,
then the user needs to change Pipe Area by changing
the value of item 305 (in sq. ft.) in Flow Basics Menu.
Subsequently user can change the engineering units
(See Menu 400). In some cases, both approaches can
be put in a single meter. For example, Channel A, B
and C can be “pipe-specific”, and Channel D can be
in velocity units. Then you can use both procedures.
However, the example below assumes four pipe-
specific ranges
For this example, I will assume the following four (4)
ranges (all Compressed Air at ~100 psig):
Channel A: 0–600 SCFM in a 2" Sch 40 Pipe
Channel B: 0–2000 SCFM in a 4" Sch 40 Pipe
Channel C: 0–5000 SCFM in a 6" Sch 40 Pipe
Channel D: 0–8500 SCFM in an 8" Sch 40 Pipe
Example 9A (Current Version):
To select a channel on Meter Rev. 2.066
(for example, Channel B) from the Keypad,
follow the steps below:
1. Verify that Meter is in Flow Mode (see Example 2)
2. Tap the Back/Down Key 4 times (in under 3
seconds)
3. Tap Flow/Shift for Channel A, Select/Set for
Channel B, Forward/Up for Channel C, or
Back/Down for Channel D
4. It will take about 10 seconds to switch channels
after step 3
Example 9B (Older Version):
To select a channel on Meter Rev. 2.065
(for example, Channel B) from the Keypad,
follow the steps below:
1. Verify that Meter is in “Flow Mode”
(see Example 2)
2. Tap the “Back/Down” Key once
3. Tap “Flow/Shift” for Channel A, “Select/Set” for
Channel B, “Forward/Up” for Channel C, or
“Back/Down” for Channel D
4. It will take about 10 seconds to switch channels
after step 3
Example 9C (Oldest Version):
To select a Channel on Meter Rev. 2.064 or
earlier (for example, Channel B) from the
Keypad, follow the steps below:
1. Press Select/Set key to enter Passcode
2. Enter 99999, then accept (Enter) to Unlock
Menuing System.
(See Example 3, if necessary.)
3. Use Flow/Shift Key to bring up System Setup Menu
(Press Flow/Shift repeatedly until you cycle to that
Menu)
4. Chan Sel Item will appear. Select it (by pressing
Select/Set)
5. Enter “1” for Channel B. (Flow/Shift moves cursor
to left to accept numeric entry)
6. Shortcut out (Hold Forward/Up key then toggle
Flow/Shift key)
7. It will take about 10 seconds to switch channels
after step 6
Example 10:
Monitoring the Sensor’s Raw Milliwatts
(RTD Power)
1. Assumes you are already in the correct Channel
(See example 9 if you have a Multiple Channel
meter)
2. Verify that meter is in Flow Mode
3. If not, refer to Example 2
4. Press Flow/Shift five times (brings up Analysis
Menu)
5. Press Forward/Up twice (brings up RTD Power)
6. Press Select/Set once
7. You are now observing the raw milliwatts (mw)
coming off the sensor circuit
8. If you want to invoke the Sensor Functionality
Zero Calibration Self Check, merely check the
observed reading (after a few minutes of “no flow”
stabilization), against the next to last line(s) of
your Meter’s Certificate of Conformance
Summary of Contents for SIE Series
Page 2: ......
Page 6: ......
Page 7: ...Section GETTING STARTED A ...
Page 8: ......
Page 23: ...Section STYLES AND FEATURES B ...
Page 24: ......
Page 28: ......
Page 29: ...Section DRAWINGS C ...
Page 30: ......
Page 39: ...Section USER INTERFACE D ...
Page 40: ......
Page 58: ......
Page 59: ...Section DIAGNOSTICS E ...
Page 60: ......
Page 63: ...Section WARRANTIES AND SERVICE WORK F ...
Page 64: ......
Page 69: ...G Section APPENDIX G ...
Page 70: ......