10
G
PREMASGARD
®
111x ⁄ 112x ⁄ 112x - SD
Rev. 2018 - V13 GB
The calibrateable compact pressure sensors of the
PREMASGARD
®
111x ⁄ 112x ⁄ 112x - SD
series are equipped with eight switchable measuring
ranges and with or without optional display (eight devices in one) and are used for measuring above-atmospheric, below-atmospheric, or differential
pressures in air. The piezo-resistive measuring element is temperature-compensated and guarantees a high degree of reliability and accuracy. These
pressure transmitters have a pushbutton for manual zero point calibration and an adjustable offset. Applications of these pressure sensors are in
clean room, medical and filter technology, in ventilation and air conditioning ducts, in spray booths, in large-scale catering facilities, for monitoring
filters, for level measurement or for triggering frequency converters. Media measured with these pressure transducers are air (non-precipitating), or
other gaseous non-aggressive, non-combustible media. The pressure sensor has eight selectable measuring ranges and therefore, minimizes the
diversity of types and inventory levels while covering a greater range of applications. The differential pressure sensor is supplied including connection
set
ASD-06
2 m connection hose, two pressure connection nipples, screws).
TECHNICAL DATA
Power supply:
24 V AC (± 20 %); 15...36 V DC for U variant
15...36 V DC for I variant,
depending on working resistance, residual ripple stabilised ± 0.3 V
Working resistance:
R
a
(ohm) = (U
b
-14 V) ⁄ 0.02 A for I variant
Load resistance:
R
L
> 5 kOhm for U variant
Power consumption:
< 1 W at 24 V DC; < 2 VA at 24 V AC
Measuring ranges:
multi-range switching with 8 switchable measuring ranges
(see table)
Output signal:
0 -10 V or 4...20 mA
Electrical connection:
2- or 3-wire connection
Media temperature:
–20...+50 °C
Pressure connection:
4 ⁄ 6 x 11 mm (hoses Ø = 4 ⁄ 6 mm), metal pressure connection nozzles
Type of pressure:
differential pressure
Medium:
clean air and other non-aggressive, non-combustible gases
Accuracy:
Type 1111 ⁄ 1121 ⁄ 1121 - SD
(1000 Pa): ± 10 Pa
Type 1115 ⁄ 1125 ⁄ 1125 - SD
(5000 Pa): ± 50 Pa
Type 1116 ⁄ 1126 ⁄ 1126 - SD
(10000 Pa): ± 50 Pa
compared to the calibrated reference device
Zero point offset:
± 10 % of final value
Above- ⁄ below-
atmospheric pressure:
max. 5 x measuring range
Long-term stability:
± 1 % per year
Signal filtering:
switchable 1 s ⁄ 10 s
Hysteresis:
0.3 % of final value
Media contacting parts:
ms, Ni, Nylon, PU, Si, PVC with plasticisers
Temperature drift values:
± 0.1 % of final value ⁄ °C
Current consumption:
< 20 mA
Linearity:
< ± 1 % of final value
Enclosure:
plastic, UV-stabilised, material polyamide, 30 % glass-globe reinforced,
colour traffic white (similar to RAL 9016)
Type 111x ⁄ 112x
:
with quick-locking screws
(slotted/Phillips head combination)
enclosure cover for display is transparent!
Type 112x - SD
:
with snap-on lid
Enclosure dimensions:
72 x 64 x 37.8 mm (Tyr 1 / Tyr 01 without display)
72 x 64 x 43.3 mm (Tyr 1 with display)
Cable gland:
M 16 x 1.5, including strain relief, exchangeable, max. inner diameter 10.4 mm
Electrical connection:
0.14 - 1.5 mm², via terminal screws
Humidity:
< 95 % r. H., non-precipitating air
Protection class:
III (according to EN 60 730)
Protection type:
Type 111x ⁄ 112x
:
IP 65
(according to EN 60 529)
Type 112x - SD
:
IP 43
(according to EN 60 529)
Standards:
CE conformity, electromagnetic compatibility according to EN 61 326, EMC directive 2014 ⁄ 30 ⁄ EU
Features: two-line
display with illumination
, cutout approx. 36 x 15 mm (W x H), to display ACTUAL pressure
ACCESSORIES
ASD - 06
Connection set (nipple straight) – (included in the scope of delivery)
ASD - 07
Connection nipple (at 90° angle)
DAL-01
Pressure outlet for ceiling or in-wall installation (e.g. in clean rooms)
WS-04