PARTICLE COUNTER / EN
09/15/17 Version No. 00
13
7-1-5 Setting the memory buffer:
The meter can store measurement data in the internal flash memory or on
micro SD card.
To set the memory buffer:
1. Press
F2
(SETUP) key on the Main screen. The Setup page 1 screen is
displayed with the current “Cnt Mode” field is selected.
2. Press
key to move to the “Buffer Select” field.
3. Use
F1
(left arrow) or
F3
(right arrow) to select “Flash” or “SD Card”.
4. Press
(RETURN) to save the choice and return to the Main
Screen.
7-1-6 Setting the alarm count and size:
When the alarm is set, an audible alarm is triggered when counts reach
the count limits on the selected particle size.
To set the alarm count and size:
1. Press
F2
(SETUP) key on the Main screen. The Setup Page 1 screen
is displayed.
2. Press
F2
(PAGE 2) key on the Setup Page 1 screen. The Setup Page
2 screen is displayed.
3. Alarm Limit is highlighted. Press
F1
(left arrow) or
F3
(right arrow)
key to move the alarm limits from zero (no alarm) to 100,000 in factors
of 10.
4. When the required limit is selected, press
key to select the “Alarm
Size” field.
5. Press
F1
(left arrow) or
F3
(right arrow) key to display the various
particle sizes.
6. Select the desired particle size and press
(RETURN) key to return
to the Main screen.
Note: When the alarm sounds, pressing
F1
(STOP) key once will mute the alarm.
Pressing
F1
(STOP) key again will stop the sample.